a3003 FC/a3004 FC
Calibration Manual
12
4-20 mA Clamp Performance Test
Do the performance test periodically and after calibration.
The performance test requires a calibrator with a current loop from the AUX terminal to the LO terminal. See Table 2.
To short the two terminals, use a conductor with 2 banana jacks at both ends. To ensure the clamp closes properly,
the diameter of the conductor should be no more than 4 mm. The Meter must be factory-calibrated to calibrate the
clamp.
To make the calibration adjustment:
1.
Make sure you have the necessary equipment, see Table 2.
2.
Allow the Meter and the 4-20 mA clamp to stabilize to room temperature.
3.
Connect the 4-20 mA clamp to a calibrated Meter.
4.
Use 1 loop insulated copper wire to short circuit the calibrator current output AUX and LO terminals, see
Figure 8.
5.
Clamp the 4-20 mA current clamp around the insulated copper wire. Current flow should be in the direction of the
arrow on the current clamp.
6.
For each value in Table 4.
a.
Zero the Meter.
b.
Apply the Calibrator Output value to the Meter.
c.
Wait 4 seconds for the signal to stabilize.
d.
Record the reading.
Table 3. a3003 and a3004 DC Meter Performance Verification
Test Connection
Calibrator Output
Resolution
Specification
UUT Reading Limits
a3003
a3004
1 Year Lower
Limit
1 Year Upper
Limit
Single-turn Loop
-100.0 mA
0.1
1.0 % ±5 digits
-101.5
-98.5
Single-turn Loop
-20.00 mA
0.01
0.2 % ±5 digits
-20.09
-19.91
Single-turn Loop
-2.00 mA
0.01
0.2 % ±5 digits
-2.05
-1.95
Single-turn Loop
2.00 mA
0.01
0.2 % ±5 digits
1.95
2.05
Single-turn Loop
20.00 mA
0.01
0.2 % ±5 digits
19.91
20.09
Single-turn Loop
100.0 mA
0.1
1.0 % ±5 digits
98.5
101.5
25-turn Coil
-80 mA (-2000 A)
1
2.5 % ±5 digits
-2055
-1945
25-turn Coil
-38 mA (-950.0 A)
0.1
2.0 % ±5 digits
-969.5
-930.5
25-turn Coil
-2 mA (-50.0 A)
0.1
2.0 % ±5 digits
-51.5
-48.5
25-turn Coil
2 mA (50.0 A)
0.1
2.0 % ±5 digits
48.5
51.5
25-turn Coil
38 mA (950.0 A)
0.1
2.0 % ±5 digits
930.5
969.5
25-turn Coil
80 mA (2000 A)
1
2.5 % ±5 digits
1945
2055
Simulate
-550 mV
1
1.0 % ±5 digits
-2533
-2467
Simulate
-190 mV
0.1
1.0 % ±5 digits
-964.5
-935.5
Simulate
-100 mV
0.1
0.2 % ±5 digits
-501.0
-499.0
Simulate
-10 mV
0.1
0.2 % ±5 digits
-50.6
-49.4
Simulate
0 mV
0.1
0.2 % ±5 digits
-0.5
0.5
Simulate
10 mV
0.1
0.2 % ±5 digits
49.4
50.6
Simulate
100 mV
0.1
0.2 % ±5 digits
499.0
501.0
Simulate
190 mV
0.1
1.0 % ±5 digits
935.5
964.5
Simulate
550 mV
1
1.0 % ±5 digits
2467
2533
Содержание A3003 FC