Wireless DC Clamp
Performance Tests
13
All readings must meet 100 % of specification, see Table 4. If the Product fails any part of the performance test,
calibration adjustments are required. See
. If the Product continues to fail, contact Fluke for repair. See
.
2000 A Clamp Performance Test
Do the performance test periodically and after calibration and the balance test.
To do the performance test:
1.
Make sure you have the required equipment. See Table 2.
2.
Connect the equipment. See Figure 6.
3.
Power on all the equipment and set the parameters:
•
On the 5522A, put in the Standby state with 0
A output.
•
On the 52120A put in the Current Mode, with the 120 A range. Put in the Local Control mode.
4.
Push
on the Meter.
5.
For each value in Table 5:
a.
Zero the Meter.
b.
Apply the Calibrator Output value to the Meter.
c.
Wait 4 seconds for the signal to stabilize.
d.
Record the reading.
All readings must meet 100 % of the specification.
Table 4. Performance Specifications for the 4-20 mA Clamp
Test
(Function)
Calibrator Outputs
Meter Response
Lower Limit
Upper Limit
mA dc
90 mA
88.6
91.4
20 mA
19.91
20.09
4 mA
3.94
4.06
-4 mA
-4.06
-3.94
-20 mA
-20.09
-19.91
-90 mA
-91.4
-88.6
Table 5. Performance Specifications for the 2000 A Clamp
Test
(Function)
List
Calibrator Output (mA)
From 5522A
Meter Response (mA)
Lower Limit
Upper Limit
A dc
1.0
0.04
0.48
1.52
50.0
2.0
48.5
51.5
300.0
12.0
293.5
306.5
950.0
38.0
930.5
969.5
1050
42.0
1019
1081
1500
60.0
1458
1543
2000
80.0
1945
2055
-A dc
-1.0
-0.04
-2.0
0.0
-50.0
-2.0
-52
-49
-300.0
-12.0
-307.0
-294
-950.0
-38.0
-970
-931
-1050
-42.0
-1081
-1019
-1500
-60.0
-1543
-1458
-2000
-80.0
-2055
-1945