ZSOQT
a.
Perform
a write
operation
to the
port
A
direction register to set
all
lines of
I/O
port
A
(PAO-PA7)
as
inputs. The write address
is 2081;
write
data
is
00.
b.
Perform
a read
operation
at the
port
A
data
register, address 2080,
while sequentially applying
both
+5V
and ground to each of the
port
A
input
pins (pins 2-9 of
U2!)
and observing the mainframe display. The
mainframe should indicate each high and
low
applied to the inputs of
port
A.
7.
Check
the output operation
of
port
B
(contained in U22 and U24) by
repeating
step
5
and using address 2083 for the
port
B
direction register,
and address 2082
for
the
port
B
data
register.
8.
Check the
input operation
of
port
B,
line PB7 (MAINTSTAT) by
repeating
step
6.
Use address 2083
for
the
port
B
direction register and
write
data
00
to set line PB7 as an input. Perform the looping read at
address 2082, and apply
both
+5V
and ground to pin
12
of the shielded
cable connector.
9.
Check
operation
of the interval timer (contained in
U25
and
26) by
first writing 00
to
address 2094 to reset the timer, then writing
OF
to
address 209C. Verify
that
the
m
output at
U20, pin
18,
goes low
in
response to the second write
operation.
10.
Check
for
the occurrence of the UUT ON signal (produced by the
timing circuits as a result of the low
lRQ
signal) at pin
19
of the
PCB—to-PCB
connector,
by repeating step
9.
ll.
Check for the occurrence of the SYNC signal at pin
10
of the
shielded cable connector, by repeating step
9.
12.
Check
both
address decoders by performing read operations at
addresses 0000, 2000, 3000, 4000, 5000, 6000, and 7000. Verify
that
the
respective decoder
output
goes low when addressed.
13.
If
repairs
have been made to the inoperative pod as a result of the
preceding checks,
reinstall
the pod’s microprocessor
and attempt
self
test. If self test operates,
but
the pod fails, refer to Troubleshooting a
Defective Pod.
The
troubleshooting
guidelines presented in the preceding
paragraphs
are
intend ed to assist
in
the
isolation
of faults within the pod.
lf attempted
troubleshooting
fails to reveal the pod fault,
return
of the pod to the nearest
Fluke Service
Center
is
recommended. Refer
to
the mainframe Service Manual
for a list of Fluke Service Centers.
5-18
Содержание 9000A-Z80QT
Страница 6: ......
Страница 14: ......
Страница 18: ...ZBOQT Figure 2 3 Connection 0 Interface Pod to UUT...
Страница 72: ......
Страница 76: ...Z8OQT MP10 H 2 9000A Z8OQT 5071 Figure 6 1 9000A 2800T Intertace Pod Final Assembly...
Страница 84: ...Z8OQT E 3 mwmdwwwwmi o a u n v u 2 as MLVLSI IQ 23333329...
Страница 85: ...ZBOQT QLVLSNXVI 9 Figure 7 1 A11 Processor PCB Assembly...
Страница 86: ...ZBOQT H H u Chum 55 ixu m3m V V v 0 A0 m v mmmcaa 5 5 taxman th XEH 35me 9 02 hww m4mm Nu W tho mu m uw m 7 4...
Страница 88: ...2800T 7 6 m V L LOl D INPUTS PAL U27 LOGIC Flgure 1 1 A11 Processor PCB Assembly cont OUTPUTS 15...
Страница 89: ...ZBOQT DEVICE 6ND PINS 14 Figure 7 1 A11 Processor PCB Assembly cont 7 7...
Страница 90: ...Z8OQT LA SELF HOLD LO BUSAK 7 8 C 2 5 5v 22A if 22L l clz Em 0531 00 463 EEF tzaF L _____________ _J...
Страница 92: ......
Страница 108: ......