Problem
Causes and Solutions
The controller controls or
attempts to control at an
inaccurate temperature
If the controller appears to operate normally except that the bath’s temperature
does not agree with the temperature measured by the user’s reference ther-
mometer to within the specified accuracy, consider the following:
Erroneous parameters.
Check that the calibration parameters are all correct
according to the Report of Test. If not, reprogram the constants. If the controller
does not keep the correct parameters, the memory backup battery may be
weak causing errors in data. See “Low Battery” in a previous solution.
Poor uniformity.
There may be an actual difference between the bath’s control
probe and the reference thermometer due to excess gradients in the bath.
Check that the bath has an adequate amount of fluid in the tank and that the
stirrer is operating properly. Also check that the reference thermometer and
control probe are both fully inserted into the bath to minimize temperature gradi-
ent errors.
Defective control probe.
Check that the control probe has not been struck,
bent, or damaged. Refer to the previous solution for how to check the probe’s
resistance.
The controller shows that
it is controlling at the
proper temperature, but
the bath temperature is
unstable
If the bath does not achieve the expected degree of temperature stability when
measured using a thermometer, consider the following:
Wrong proportional band setting.
If the proportional band is set too narrow,
the bath will oscillate causing poor stability. In this case, increase the width of
the proportional band.
If the proportional band setting is too wide, the long-term stability of the bath is
affected. In this case decrease the width of the band. (Refer to Section 9.7.)
Bath fluid is too thick.
Make sure that the bath fluid used is less than 50 centi-
Stokes (10 is ideal) at the temperature at which the bath is controlling. Check
the fluid manufacturer’s specifications.
You should also change the bath fluid regularly and if it changes colors or be-
comes too thick.
Defective control probe.
Check that the control probe has not been struck,
bent, or damaged. Refer to the previous solution for how to check the probe’s
resistance.
The controller alternately
heats for a while then
cools
Wrong proportional band setting.
If the proportional band is set too narrow,
the bath will oscillate between too much heating and too much cooling causing
instability. Increase the width of the proportional band until the temperature sta-
bilizes. (Refer to Section 9.7.)
The bath does not
achieve low temperatures
Insufficient cooling
. This may be caused by lack of refrigerant because of a
leak in the system. Refer to Section 12, Charging Instructions.
77
14 Troubleshooting
Troubleshooting