data, denoted by “n”, may be entered in decimal or exponential notation. Char-
acters are shown in lower case although upper case may be used. Spaces may
be added within command strings and are simply ignored. Backspace (BS,
ASCII 8) may be used to erase the previous character. A terminating CR is im-
plied with all commands.
10.4
Power Control Functions
The digital interface is capable of controlling the cooling function so that the
instrument can be remotely operate at any temperature within the range of the
bath. To allow the interface to control the cooling, disable the front panel con-
trols by switching the refrigeration switch to OFF. Otherwise, the interface
would not be able to switch these functions off.
To control the refrigeration power with the digital interface, the front panel re-
frigeration switch must be off. The refrigeration power function is controlled
with the “F2” command. Setting the “F2” value to 0 turns the refrigeration off
and setting the “F2” value to 1 turns the refrigeration on. “F2” alone returns 0
or 1 showing the state of the refrigeration power control as shown in the fol-
lowing table.
Function
Command
0
1
Refrigeration
F2
off
on
63
10 Digital Communication Interface
Power Control Functions
Содержание 7312