U
Increment display
C 30.00
New set-point value
Press the “SET” button to accept the new value and to display the vernier value.
The bath begins heating or cooling to the new set-point.
S
Store new set-point, access vernier
0.00000
Current vernier value
Press the “EXIT” button and the bath temperature is displayed again.
E
Return to the temperature display
24.73 C
Bath temperature display
The bath heats or cools until it reaches the new set-point temperature. Turn off
the cooling to reach and control at higher temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cutout should be correctly set for
added safety. See Section 9.8, Cutout.
To obtain optimum control stability adjust the proportional band as discussed in
Section 9.7, Proportional Band.