12
Charging Instructions
The 7312 uses R-134a with a polyolester oil. Care must be taken to avoid con-
tamination from other types of refrigerants and oils.
12.1
Leak Testing
Leak testing should be done with equipment designed for use with R-134a.
Bubble, electronic halogen detector, or ultra-sonic leak testing may be viable in
some instances.
12.2
Evacuation
DO NOT leave the system open for more than 15 minutes. Polyolester oils are
very hygroscopic. Evacuate the system to a minimum of 200 microns. Evacuate
from both high and low sides of the system. Schrader valves provide access to
the system.
12.3
Charging
After evacuation, charge the system with 160 grams of R-134a. Set the suction
pressure to 8-10 psig for operation.
NOTE:
A fan must be placed in front of the condenser if the unit is oper-
ated with the side panels removed. The unit requires the side panels to be
installed for proper airflow across the condenser.
71
12 Charging Instructions
Leak Testing