Most probes have handle temperature limits. Be sure that the probe handle
temperature limit is not exceeded in the air above the instrument.
•
The instrument and any thermometer probes used with it are sensitive in-
struments that can be easily damaged. Always handle these devices with
care. Do not allow them to be dropped, struck, stressed, or overheated.
COLD BATHS
•
Refrigerated baths require that the condensing coil be cleaned periodi-
cally. Accumulation of dust and dirt on the condenser will result in pre-
mature failure of the compressor.
•
This bath has been equipped with a brownout and over voltage protection
device as a safety feature to protect the system components.
•
Mode of Operation:
This bath needs to be plugged into the line voltage
for at least 2 minutes before operation. This is only necessary for the first
time that the bath is energized or when it is moved from one location to
another. Turning the bath ON or OFF does not trigger the delay.
•
If a High/Low voltage condition exists for longer than 5 seconds, the bath
de-energizes. An amber indicator on the back panel lights when this con-
dition exists.
•
Re-energization is automatic upon correction of the fault condition and af-
ter a delay cycle of about 2 minutes. If a fault condition exists upon appli-
cation of power, the bath will not energize.
•
Under and Over Voltage Protection at 115 VAC
♦
Voltage Cutout: ±12.5% (101 - 129 VAC)
♦
Voltage Cut In: ±7.5% (106 - 124 VAC)
•
Under and Over Voltage Protection at 230 VAC
♦
Voltage Cutout: ±12.5% (203 - 257 VAC)
♦
Voltage Cut In: ±7.5% (213 - 247 VAC)
1.3
Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010