ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
LF
Serial linefeed parameter
Press the “SET” button to access the linefeed parameter.
LF=On
Current linefeed setting
The mode may be changed using the “UP” or “DOWN” button and then press-
ing the “SET” button.
LF=OFF
New linefeed setting
9.13
IEEE-488 Parameters
Baths may optionally be fitted with an IEEE-488 GPIB interface. In this case
the user may set the interface address and the transmission termination charac-
ter within the IEEE-488 parameter menu. This menu does not appear on baths
not fitted with the interface. The menu is indicated by,
IEEE
IEEE-488 parameters menu
Press the “SET” button to enter the menu.
9.13.1
IEEE-488 Address
IEEE-488 interface must be configured to use the same address as the external
communicating device. The address is indicated by,
AddrESS
IEEE-488 interface address
Press the “SET” button to access the address setting.
Add=22
Current IEEE-488 interface address
Adjust the value with the “UP” or “DOWN” button and then use the “SET”
button to set the address to the displayed value.
Add=15
New IEEE-488 interface address
9.13.2
Transmission Termination
The transmission termination character can be set to carriage return only, line-
feed only, or carriage return and linefeed. Regardless of the option selected, the
7312 TPW Maintenance Bath
User’s Guide
54