![Fluke 7312 Скачать руководство пользователя страница 53](http://html.mh-extra.com/html/fluke/7312/7312_user-manual_2309393053.webp)
Par
Operating parameters menu
Press the “UP” button to enter the menu. The operating parameters menu con-
tains the cutout reset mode parameter.
9.11.1
Cutout Reset Mode
The cutout reset mode determines whether the cutout resets automatically when
the bath temperature drops to a safe value or must be manually reset by the
operator.
The parameter is indicated by,
CtorSt
Cutout reset mode parameter
Press the “SET” button to access the parameter setting. Normally the cutout is
set for manual mode.
Cto=rst
Cutout set for manual reset
To change to automatic reset mode, press the “UP” button and then the “SET”
button.
Cto=Auto
Cutout set for automatic reset
9.11.2
Triple Point of Water Cutout
The triple point of water cutout protects the TPW cell in case of system failure,
which may cause the bath to rapidly cool below the maintenance set-point. This
could cause the cell to rupture due to the expansion of ice. The cutout uses a
thermistor probe sensor and disengages the refrigeration compressor. If the
probe is replaced, the cutout board must be recalibrated. (See Section 12). The
user may set the cutout temperature. A typical set-point is from –0.3 to –0.5
°
C.
If undesired trips occur, the temperature may be reduced. The adjustment range
is
±
1.27
°
C from 0
°
C in 0.01
°
C increments.
The triple point of water cutout engages for three reasons: 1) the controller
does not send a watch dog signal to the cutout board (controller lockup), 2) the
thermistor sensor is disconnected or opens, or 3) the bath goes below the cutout
set-point. (This could be if the temperature controller set-point is too low or if
the heater fails.) When the TPW cutout is triggered, the display alternately
flashes tP-cto and the current bath temperature. The serial port outputs “TPW
Cutout” every three seconds. Pressing any button causes the cutout error mes-
sage to stop being output until another TPW cutout error has occurred. Each
time the cutout is tripped, the bath temperature set-point automatically in-
creases in 0.5
°
C increments until the set-point has reached a maximum of
25
°
C.
The parameter is indicated by,
51
9 Controller Operation
Operating Parameters