![Fluid Imaging Technologies FlowCAM Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/fluid-imaging-technologies/flowcam/flowcam_manual_2308799076.webp)
- 76 -
Show Efficiency
: Displays the actual efficiency (available only if a syringe pump is used).
Setup and Focus Mode… Tools Menu
Recalibrate:
This function is not relevant when in Setup and Focus Mode. This function is employed
when doing an analysis. During the analysis, the user can select the Recalibrate function on this menu to
immediately generate a new calibration image.
Reset Frame Rate Counter:
This function will reset the frame rate counter (if displayed) to zero. The
image count and elapsed run time, if displayed, are also reset.
Raw Pixel Intensity…
This advanced tool can be used as a guide to help set the illumination uniformity of the image when
adjusting either the LED adjustment knob or the camera Gain settings. A Raw Image Pixel Intensity
graph is shown that has both a histogram of pixel intensity and the cumulative % for the histogram. The
intensity at the lower right hand end of the histogram line shows the maximum intensity. If the intensity
is not optimally adjusted, the graph may appear similar to the following example:
Notice the uneven appearance of the histogram line (red). As the intensity is adjusted, the graph should
change to look similar to the following example: