Attach waste tubing to pump
§
Insure that waste tubing is going to a waste container
§
Center the fine focus travel if required
5.
Setup the software/hardware in the following order:
§
Open Visual Spreadsheet and select the correct objective
§
Insure that the Status Bar color (if displayed) at the bottom of the Main Window matches the
colored stripe on the installed objective
§
Main Window: Setup > Setup and Focus
·
Focus on the flow cell using the coarse adjustment Z translator
·
Use the Course Adjustment screw on the Z translator to lock the Z translator into place
·
Add the sample
and turn on the pump until the sample can be seen in the Setup and Focus
window. Note: Proper sample preparation is critical to having a successful FlowCAM
analysis. It is important to get the proper particle size distribution and sample dilution to
ensure that the best possible data are obtained.
·
Use the fine focus adjustment knob to adjust the focus until the sample appears to look
the best possible
·
Setup and Focus window: Setup > Camera
·
Move the Camera Settings window so that it is not overlaying the Setup and Focus
window (Note: You must click “Apply” for any changes in the Camera window to take
effect).
¨
Adjust the U and V color balance until the displayed camera image is a uniform gray
background (only necessary for a color camera)
¨
Adjust the Gain and Flash Duration (increased gain and flash duration will result in a
higher intensity mean) until you have an intensity mean of 165-185
¨
Close the Camera window
·
Determine the concentration, coloration, and size distribution of the sample by looking at
the particles that are displayed in the setup and focus window
§
Close the Setup and Focus window
§
Main Window: Setup > Context
·
Context Window: Adjust the context settings via the following tabs:
¨
Capture
¨
Flow Cell