Step 1: Dismounting of the transducer mounting fixture Variofix L
• Disassemble the transducer mounting fixture Variofix L, see Fig. 6.13.
Step 2: Fixing the clasps to the tension straps
• Select the installation instructions according to the supplied clasp:
Band clamp clasp
The clasp is fixed to the tension strap, see Fig. 6.14.
Quick release clasp
The clasp is fixed to the tension strap, see Fig. 6.15.
• Cut the tension strap to length (pipe circumf at least 120 mm).
Ratchet clasp
• Cut the tension strap to length (pipe circumf at least 120 mm).
Fig. 6.13:
Dismounting of the transducer mounting fixture Variofix L
1 – cover
2 – screw
3 – nut
4 – rail
5 – tension strap clamp
Fig. 6.14:
Band clamp clasp with tension strap
Fig. 6.15:
Quick release clasp with tension strap
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
→
Debur sharp edges.
→
Wear the required personal protective equipment. Observe the applicable rules.
1
2
3
4
5
5