16 Data transmission
FLUXUS WD
16.1 Service interface
123
UMFLUXUS_WDV1-3EN, 2018-07-16
16
Data transmission
The data is transmit to the transmitter via service or process interfaces (optional).
16.1
Service interface
By means of the program FluxDiagReader it is possible to transmit via service interfaces (USB, Ethernet) data to the PC.
For the connection of the service interfaces see section 7.4.
The following tasks can be carried out:
• read and store measured values, setup settings and snaps
• graphically display measured value
• export of data in csv format
For the operation of the program see FluxDiagReader support.
16.1.1
Ethernet interface
In order to use the Ethernet interface it is important to enter the network parameters.
• Select the program branch
Special functions
.
• Press ENTER.
• Select the menu item
Network
in the program branch
Special functions\Communication
.
• Press ENTER.
Manual input
• Select
Manual
to enter the network parameter (IP address, subnet mask and standard gateway).
default value in the transmitter:
– IP address: 192.168.0.70
– subnet mask: 255.255.255.0
– standard gateway: 192.168.0.1
Automatic addressing with DHCP
• Select
Automatic
to automatically identify the network parameter (IP address, subnet mask and standard gateway) via
a DHCP server.
• Select the menu item
Show auto config.
in the program branch
Special functions\Communication\
Network
to display the identified network parameters.
• Press ENTER.
Special functions\Communication\Network
Notice!
The entered network parameter has to accord with the network parameters of the LAN network.
Notice!
The network parameters can only be automatically identified if the LAN network supports DCHP.
Notice!
The program FluxDiag is used to transmit data from the PC to the transmitter.
Содержание FLUXUS WD
Страница 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Страница 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Страница 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Страница 171: ...Annex FLUXUS WD D Conformity declarations 171 UMFLUXUS_WDV1 3EN 2018 07 16 D Conformity declarations...
Страница 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Страница 173: ......
Страница 174: ......