17 Advanced functions
FLUXUS WD
17.4 Diagnosis by means of the snap function
133
UMFLUXUS_WDV1-3EN, 2018-07-16
• Select the source item
Extended diagnost.
• Press ENTER.
• Select the list item for which a condition is to be defined, see Tab. 17.3.
• Press ENTER.
17.4
Diagnosis by means of the snap function
17.4.1
Configuration
By means of the snap function it is possible to store measuring parameters which are useful for the evaluation of measur-
ing results or diagnosis purposes. To make use of the snap function it has to be configured.
• Select the menu item
Configuration
in the program branch
Special functions\Snap
.
• Press ENTER.
• Select
On
to activate the snap function.
• Press ENTER.
• Select
Yes
to activate the snap ringbuffer.
If the snap ringbuffer is activated, after taking the 51th, snap the oldest snaps are overwritten. If the ringbuffer is deactivat-
ed, up to 50 snaps can be stored.
• Press ENTER.
• Select
Yes
to activate
Auto snap
.
If
Auto snap
is activated, snaps are automatically stored during a measurement failure.
• Press ENTER.
• Select
Yes
if an event has been parameterized on the event trigger R1 which, at time of occurrence, has to release a
snap.
• Press ENTER.
Options\Channel Y\Event trigger\Enable Rx\Source item
Special functions\Snap\Configuration
Special functions\Snap\Configuration\Snap
Special functions\Snap\Configuration\Snap ringbuffer
Special functions\Snap\Configuration\Auto snap
Special functions\Snap\Configuration\Snap on R1
Содержание FLUXUS WD
Страница 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Страница 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Страница 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Страница 171: ...Annex FLUXUS WD D Conformity declarations 171 UMFLUXUS_WDV1 3EN 2018 07 16 D Conformity declarations...
Страница 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Страница 173: ......
Страница 174: ......