Annex
A Menu structure
FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
164
Example 2
The flow velocity is measured.
If the flow velocity is
≤
5 m/s, the transmitter measures in the TransitTime mode. As long as the flow velocity is > 5 m/s, the transmit-
ter has to measure in the FastFood mode.
trigger:
condition:
action:
flow velocity > 5 m/s
R1 with flow velocity > 5 m/s
measurement in the FastFood mode
Options
Options
Channel A
Channel A
Event trigger
Remote functions
Select R1
Enable FastFood
→
R1
Enable R1
Yes
Source item
→
Flow quantities
Flow quantities
→
Flow velocity
Function
→
MAX (x>limit)
Type
→
Non-hold
Trigger value
→
5 m/s
Hysteresis
→
0.4 m/s
Glitch interval
→
1.0 s
Failure delay
→
1.0 s
1
2