15 Data logger
FLUXUS WD
15.2 Deletion of the data logger
121
UMFLUXUS_WDV1-3EN, 2018-07-16
Acoustic signal during the storing
The storing of each measured value can be acoustically signalized. The signal can be activated or deactivated in the pro-
gram branch
Special functions\
Data logger
\Configuration\Beep on storage
.
• Select
On
to activate the acoustic signal. Select
Off
to deactivate the acoustic signal.
• Press ENTER.
15.2
Deletion of the data logger
• Select the menu item
Delete meas. values
in the program branch
Special functions\Data logger
.
• Press ENTER.
• Select
Yes
or
No
.
• Press ENTER.
15.3
Information relating the data logger
• Select the menu item
Data logger info
in the program branch
Special functions\Data logger
.
• Press ENTER.
The following information relating the data logger is displayed:
15.4
Print of measured values
• Start the terminal program.
• Enter the transmission parameters into the terminal program. The transmission parameters of the terminal program and
the transmitter have to be identical, see section 15.5.
• Select
Special functions\Data logger\Print meas. values
.
This display will only be indicated if the transmitter has an RS485 interface.
• Press ENTER.
Special functions\Data logger\Configuration\Beep on storage
Special functions\Data logger\Delete meas. values
Special functions\Data logger\Data logger info
display
description
Activated:
activated/deactivated data logger
This display will only be indicated if the measurement has started and the data logger is activated.
Logger full:
data logger is full/not full
Ringbuffer:
ringbuffer is activated/deactivated
Meas. val. series:
number of stored series of measured values
Occup. logger.:
percentage of memory actually used
Special functions\Data logger\Print meas. values
Содержание FLUXUS WD
Страница 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Страница 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Страница 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Страница 171: ...Annex FLUXUS WD D Conformity declarations 171 UMFLUXUS_WDV1 3EN 2018 07 16 D Conformity declarations...
Страница 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Страница 173: ......
Страница 174: ......