3 General principles
FLUXUS F60*
3.3 Acoustic penetration
23
UMFLUXUS_F60xV5-0EN, 2017-10-16
Vertical pipe
Select the measuring point at a pipe location where the fluid flows upward. The pipe has
to be completely filled, see Fig. 3.12 and Fig. 3.13.
Free inlet or outlet pipe section
Select the measuring point at a pipe section where the pipe cannot run empty, see Fig.
3.14 and Fig. 3.15.
Fig. 3.12:
Recommended transducer
mounting position
Fig. 3.13:
Disadvantageous transducer
mounting position
Fig. 3.14:
Recommended transducer
mounting position
Fig. 3.15:
Disadvantageous transducer
mounting position