19 Settings
FLUXUS F60*
19.2 Measurement settings
189
UMFLUXUS_F60xV5-0EN, 2017-10-16
19.2
Measurement settings
• Select
Special Funct.\SYSTEM settings\Measuring
.
• Press ENTER.
• Select
on
to enable the NoiseTrek mode,
off
to disable it.
• Press ENTER.
• Select
yes
if the measured sound speed is to be compared to the theoretical or expect-
ed value.
• Press ENTER.
The difference
δ
c = c
mea
- c
stored
between the two sound speeds will be displayed in the
lower line during the measurement. c
stored
is the sound speed of the reference fluid
stored in the database.
• Press key
during the measurement to scroll to the display of
δ
c.
• Select
normal
to display and transmit the profile corrected flow values,
uncorr.
to
display and output the flow values without flow profile correction.
• Press ENTER.
For further information see section 17.6.
An upper limit for the flow velocity can be entered, see chapter, section 17.4.
• Enter zero to deactivate the flow velocity check.
• Press ENTER.
Special Funct.\SYSTEM settings\Measuring
Notice!
The settings will be stored at the end of the dialog. If the menu item is quit by press-
ing key BRK, the changes will not be stored.
Special Funct.\...\Measuring\Enable NoiseTrek
Special Funct.\...\Measuring\Compare c-fluid
Special Funct.\...\Measuring\Flow Velocity
Special Funct.\...\Measuring\Velocity limit
Содержание FLUXUS F60 Series
Страница 1: ...FLUXUS F60 UMFLUXUS_F60xV5 0EN Operating instruction...
Страница 218: ...21 Heat flow measurement 21 6 Steam in inlet FLUXUS F60 2017 10 16 UMFLUXUSF60xV5 0EN 218...
Страница 241: ...Annex FLUXUS F60 D Conformity declarations 241 UMFLUXUS_F60xV5 0EN 2017 10 16 D Conformity declarations...
Страница 242: ...Annex D Conformity declarations FLUXUS F60 2017 10 16 UMFLUXUSF60xV5 0EN 242...
Страница 243: ......
Страница 244: ......