25 ft clear tubing (
⅜
” (9.5 mm) O.D.) for
FLEX OUT
connection
25 ft large blue tubing (
⅜
” (9.5 mm) O.D.) for vacuum source connection
Flexcell
FX-5000
Tension System User's manual, Culture Plate & Loading Stations
™
manual, Vacuum
Source Connection Manual, Water Traps and Drying Filter Tech Report and other ancillary manuals for
computer and monitor
Surge protected power outlet strip
T
ISSUE
T
RAIN
®
A
CCESSORY
K
IT
(O
PTIONAL
)
(Note: If purchased as a system, the Tissue Train
®
Accessory Kit is substituted for the Flexcell
®
vacuum pressure
baseplate, gaskets, and acrylic window for BioFlex
®
culture plates and BioFlex
®
Loading Stations
™
)
Flexcell
®
vacuum pressure baseplate, gaskets, and acrylic window for Tissue Train
®
culture plates
Arctangle
®
Loading Stations
™
and Lubricant
Trough Loader
™
Loading Stations
FX-5000
™
T
ENSION
C
OMPUTER
S
PECIFICATIONS
Dell OptiPlex Small Form Factor
Intel Core i5 processor (3.2 GHz)
250 GB SATA 6.0 Gb/s hard drive
DVD +/- RW Drive
2 GB of DDR3 RAM
Integrated video, Intel
®
HD2500
Operating System:
Windows 7 Professional
FX5K
™
T
ENSION
F
LEX
L
INK
C
OMPONENT
S
PECIFICATIONS
Custom design vacuum controller board
o
Microprocessor sampling rate for display data transducer: 200 Hz
o
Microprocessor sampling rate for feedback/error (valve adjustment) transducer: 1 KHz
2 proportional-solenoid valves
Ethernet port (RJ45)
FX5K
™
T
ENSION
F
LEX
L
INK
®
C
ALIBRATION
S
PECIFICATIONS
Unconstrained Static Performance (low side transducers):
Maximum allowable waveform variance from static line: ± 0.25% elongation
Maximum % elongation error (actual vs. programmed): +/-0.6% elongation
Maximum allowable transducer inaccuracy: +/-1.0 kPa
Unconstrained Dynamic Performance (low side transducers):
Maximum % elongation error (actual vs. programmed): +/- 0.5% elongation, at the following frequencies:
<5% elongation, 0.5 and 1.0 Hz; 10 % elongation, 0.5 and 0.75Hz; >10% elongation, 0.5 Hz
Maximum allowable transducer inaccuracy: +/-1.0 kPa
BioFlex 25mm Loading Station Static Performance (high side transducers):
Maximum allowable waveform variance from static line: +/- 0.25% elongation
Maximum % elongation error (actual vs. programmed): +/-0.5% elongation
Maximum allowable transducer inaccuracy: +/-2.0 kPa
BioFlex 25mm Loading Station Dynamic Performance (high side transducers):
Maximum % elongation error (actual vs. programmed): +/-0.75% elongation, at
the following frequencies: <15% elongation, 0.5 and 1.0 Hz; >17.5% elongation, 0.5Hz
Maximum allowable transducer inaccuracy: +/-2.0 kPa