flexcell FX-5000 Скачать руководство пользователя страница 25

 

F

LEXCELL

®

 

I

NTERNATIONAL 

C

ORPORATION

 

 

 

19

instructions on programming additional steps. 
 

Problem 5: Plate wells are drying up during 
the experiment 

 

Solution 5: 

  When vacuum is pulled to expand the 
silicone membranes at the bottom of the 
BioFlex

®

 wells, the volume of the well 

increases, pulling air into the well from the 
atmosphere. When the membranes contract, 
the air is pushed back out. Because of this 
cyclic airflow during any experiment, the 
fluid evaporation rate in the BioFlex

®

 wells 

will increase. The best solution to this 
problem is to use more culture medium.  
 

Problem 6: “An invalid argument was 
encountered" error is shown when 
attempting to assign a regimen. 

 

Solution 6: 

 

This problem is caused when a user name 

is entered into the software without any 
regimens listed under his/her name. If the 
assign function sees that a user has nothing to 
assign an error is displayed. Simply delete the 
user(s) without a regimen or make and save a 
regimen under the user(s) name(s). 
 

Problem 7: Water is accumulating quickly 
in the Water Traps 

 

Solution 7:   

This problem is likely due to a poor gasket 

seal on the BioFlex

®

 baseplate. See the 

solution to Problem 1 to correct this. Be sure 
to empty the Water Traps frequently if this 
problem occurs. Water accumulation in the 
FlexLink

®

 can damage the valves. 

 

Problem 8: Cells are growing only in the 
center of the wells due to a) excessive 
volume of medium in the well, causing 
pooling of the cells in the middle of the well 
due to gravity or b) incomplete 
trypsinization or over-trypsinization of 

cells, causing clumping and uneven 
distribution of cells 

 

Solution 8: 

a)

 

Use 2-3 ml of media at plating. After 
cell adhesion, add medium to 5 ml. 
Start a flexing regimen prior to cell 
attachment; cells will be distributed 
evenly over the growth surface during 
active flexing. 

b)

 

Trypsinize once for about 10-15 
minutes, or less, depending on the 
cell type being used. 

 

Problem 9: Cells are not attaching 
properly to the growth surface or are 
detaching during stretching 

 

Solution 9: 

 

This problem can be caused by plating 

too many cells. Try plating with a lower cell 
density—if the cells are not confluent, they 
will form a matrix with the collagen (or other 
matrix coating), attaching more securely to 
the growth surface. Also, check with 
Flexcell

®

 to be sure that the best matrix 

coating for attachment of the particular cell 
type is being used. 

Problem 10: One or more of the FX5K

 

Tension FlexLinks

®

 connected to the 

system are not being recognized by the 
software. This is usually represented by 
the error message “Cannot Connect to 
Controller” when attempting to connect a 
FlexLink

®

 to assign a regimen. 

 

Solution 10:   

 

This problem is usually due to a lack of 

communication between the computer and 
one or more FX5K

 Tension FlexLinks

®

Check the FX5K

 Tension FlexLink

®

 

Ethernet port connection on the computer 
and the communications connection on the 
back of the FlexLink

®

 to be sure that these 

are secure. Check the FlexLink

®

 to see if the 

power is ON (the FlexLink screen should 
power on). If both of these conditions are 

Содержание FX-5000

Страница 1: ...21 0 Culturing Cells in a Mechanically Active Environment Flexcell International Corporation 2730 Tucker Street Suite 200 Burlington NC 27215 800 728 3714 919 732 1591 FAX 919 732 5196 www flexcellin...

Страница 2: ...uctions 7 3 1 Starting the FX 5000 Program 7 3 2 Using the FX 5000 Tutorial 7 3 3 Main Display Window 7 3 4 Configuring Users 9 3 5 Creating Regimens 10 3 6 Creating a Regimen with a Custom Waveform S...

Страница 3: ...than our factory or distributor Breaking the seal and opening the unit case or violating unit components will void the warranty Failure to run the drying regimen as recommended in this manual will vo...

Страница 4: ...nty are listed below You will need to have the service tag number and Express Service Code number which are located on the side of the computer Dell Hardware Limited Warranty Plus Onsite Service Exten...

Страница 5: ...to a properly grounded power outlet and used with the power cord supplied with the unit s unmodified 3 The unit must be operated with the factory installed cover in place 4 No modifications of the equ...

Страница 6: ...seplate s in the incubator a Insert the 6 4 mm and 9 5 mm tubing into the access opening of the incubator b Connect the quick disconnects for the 6 4 mm and 9 5 mm tubing to the baseplate s c Place fo...

Страница 7: ...D for vacuum source connection 16 Flexcell FX 5000 Tension System user s manual and other ancillary manuals for computer and monitor 17 Surge protected outlet strip 18 Water Trap x2 19 Vacuum source o...

Страница 8: ...the system is 48 37 lbs 21 96 kg Figure 3 Dell OptiPlex rear view 2 3 SETUP OF THE COMPUTER AND FX5K TENSION FLEXLINK Setup the Dell OptiPlex desktop computer and peripherals according to the manufac...

Страница 9: ...onto the fitting Tighten each nut by hand until the tubing is crimped to minimize vacuum leakage Connect the power cables 115 VAC or 230 VAC to the rear power inlet FlexLink Power of the FlexLink Fig...

Страница 10: ...ure that the gaskets are fully pressed onto the culture plates and that no parts are folded under 4 Place the culture plates and gaskets into the baseplate wells Check to be sure that the gaskets are...

Страница 11: ...use the auto control pump feature plug the vacuum pump power cable into the PUMP CONTROL CONNECTION outlet on the rear of the FX5K Tension FlexLink Fig 2 and 7 NOTE When using the auto pump control fe...

Страница 12: ...imen Ready User name Platform Regimen Name Flexcell International Corp www flexcellint com Regimen is running or being simulated 1 Regimen Left 5 min User name Platform Regimen Name Flexcell Internati...

Страница 13: ...ength as much as possible between the vacuum source and the baseplate See page 16 for more information on obtaining the optimal performance of your FX 5000 Tension System 3 1 STARTING THE FX 5000 PROG...

Страница 14: ...al data patterns may be confounded with noise For the case of the pressure plot as a function of voltage from the FX 5000 Tension System pressure transducer the rough data yield a plot with a coeffici...

Страница 15: ...changed by clicking on one of the two small double arrow buttons at either end of the axis The button with arrows pointing away from each other increases the axis scale while the button with arrows po...

Страница 16: ...boxes under Min and Max 4 Enter the frequency for the waveform 0 01 5 Hz in the box under Freq 5 If a static regimen is required enter the desired elongation at which the membrane is to be held in the...

Страница 17: ...ed regimen name in the Regimens for Selected User box and then click on Delete Regimen Confirm the deletion by clicking on Yes NOTE If the regimen to be deleted is currently assigned the regimen must...

Страница 18: ...ll be MAXIMUM and the regimen cannot be saved Adjust the time and or cycle number to decrease the total duration time 7 Click on Save Regimen 8 When finished with regimen editing click on Close 3 7 AS...

Страница 19: ...unusual elements is to accommodate for the lag time during sudden changes in pressure e g rising falling edge of a square wave Without these unusual elements the valve controllers may oscillate for a...

Страница 20: ...l turn the pump ON The software will turn the pump OFF when a regimen is complete This option also allows the pump to be turned OFF during a regimen step when a static waveform regimen is selected tha...

Страница 21: ...box can be set so that the data viewed in the plot begin at the specified time entered here Use the timeline at the top of the graph to navigate through the regimen Users can move the slider function...

Страница 22: ...the incubator and stored flat to prevent warping Do not hang the red gaskets on pegs to dry They will stretch over time 3 18 EXITING THE FX 5000 PROGRAM To exit the FX 5000 software in the Controller...

Страница 23: ...elongation and frequency a Shorten the FLEX IN FLEX OUT and SYSTEM tubing to their minimum possible lengths See section 2 2 for minimum lengths b Use a more powerful vacuum source if recommended spec...

Страница 24: ...Regimen wave is flatlining Solution 2 Remove the SYSTEM vacuum line at the back of the FlexLink and check to be sure that there is vacuum coming through the line Check the VENT port at the back of the...

Страница 25: ...of the cells in the middle of the well due to gravity or b incomplete trypsinization or over trypsinization of cells causing clumping and uneven distribution of cells Solution 8 a Use 2 3 ml of media...

Страница 26: ...y dry before connecting to the FX5K Tension FlexLink Contact time time surface remains wet of the above solutions is critical to the effectiveness of disinfecting Times given are minimal and longer co...

Страница 27: ...re from min to max Triangle Wave This waveform provides a linear ramped rise to the max and a linear ramped fall back to the min Square Wave This waveform provides the fastest possible rise to max fol...

Страница 28: ...inusoidal wave Heart wave Electronic and Pressure Triangular wave Square wave and custom waveforms Provides equibiaxial strain or unconstrained distension to the BioFlex culture plate membrane Optiona...

Страница 29: ...m design vacuum controller board o Microprocessor sampling rate for display data transducer 200 Hz o Microprocessor sampling rate for feedback error valve adjustment transducer 1 KHz 2 proportional so...

Страница 30: ...equibiaxial strain Minimum achievable elongation with the FX 5000 1 2 Maximum achievable elongation with the FX 5000 8 0 FX 5000 TENSION PHYSICAL SPECIFICATIONS FX 5000 Computer Size W x H x D 3 65 x...

Страница 31: ...ion 20 Frequency 1 Hz The maximum elongation achieved might only be 10 Reducing the frequency of the regimen to 0 5 Hz will produce an achievable elongation of 18 The maximum allowable elongation for...

Отзывы: