Section 7
Trouble Shooting
Trouble During
Operation
Causes
Solution
Probe lamp is not ON
Probe connection failure.
Reconnect probe in right way.
Damage due to chemical
spill or overheat.
Contact Fisher Scientific for service.
Error Messages
If
you see below “Error codes”, turn off the power. Restart after having the
unit rest enough.
Error Code
Cause
Effect
ERRO NO 1
Related to
Heater system
If Top plate is
overheated
Heater off
ERRO NO 2
If sensor detects a
sudden temperature
changes
Heater off
ERRO NO 3
If PCB is overheated
Heater off
ERRO NO 4
If sensor is not
connected correctly
Heater off
ERRO NO 5
If temperature difference
has arisen between two
sensors
Heater off
ERRO NO 6
Related to
temperature
probe
If temperature probe is
in the air
Heater off
ERRO NO 7
Temperature probe is
not soaked enough into
media
Heater off