2-1
Section 2
Functional Description
Features
• Offset
To use your own thermometer for temperature control in specific
applications, there can be some differences between the temperature
of your thermometer and the displayed temperature of this unit. If
needed, you can offset such temperature differences ranging from -
10°C to ±50°C at 0.1°C intervals.
• Selection of the Timer Mode
Immediate Activation (T1): The timer starts immediately after setting
the timer, Delayed Activation (T2): The timer is activated only when
the set temperature is reached.
• Quick Heat Up Time
Ample heating capacity and superb heat transfer rate of the tightly
integrated structure of heater and ceramic-coated aluminum alloy top
plate allow quick heat up time.
• The ceramic-coated top plate is highly resistant to heat and corrosion.
In addition, its white color is optimal for monitoring color changes of
the media during operation.
Safety
• Hot Top Warning Indicator
The top plate temperature can remain very hot for some time even
after the heater is turned off. To prevent injury or fire under such
circumstances, this instrument has a hot top warning indicator on the
control panel. This indicator will illuminate if the top plate temperature
is over 50°C. Even so, do not rely on this indicator alone for your
safety
• Overheating prevention device
Basically built in electronic overheating prevent device and
mechanical body overheating prevent device.
• Heating plate overheating prevention device
A heating plate overheating prevention device is particularly built-in.
• The devices overheat warning function
The devices WARNING LED will be lighted. In case of the devices
temperature goes over safe range.