Temperature Probe
Control Mode
For temperature and timer setting on the Temperature Probe Control
Mode, see Temperature Setting and Timer Setting.
In case of low smoke point media heating, High temperature limit value
should be lower than smoke point. [Refer to High Temperature Limit
Setting]
(1) Temperature Probe Errors
On temperature control mode if the external probe is in the air or is not
enough immersed into the media, error message will be shown on
display and heater will be automatically turned off.
For correct control media by the external temperature probe, make sure
the sensor to completely immerse into the media (more than 20mm
deep). If the sensor is in the air due to evaporation of media, precise
temperature control may fails.
[Refer to Error Message]
Error code
Cause
Effect
ERRO NO 6
Temperature
Probe Errors
If the temperature probe
is in the air
Heater off
ERRO NO 7
If the temperature probe
is not completely immer
sed into the media
Heater off
• Hot Top Warning Indicator will illuminate if top plate temperature is
over 50°C.
• For correct control media by the temperature probe, make sure the
sensor to completely immerse into the media (more than 20mm
deep). If the sensor is in the air due to evaporation of media, Errors
will be shown on display.
• Probe operating temperature range is :
B Class : Max 250°C.
A Class : Max 400°C
4-13
Section 4
Control Panel