Maximum Heating Rate 

Setting (LIMt)

 

Step  4:  Save  the  desired  limit  by  using  the  Stirrer  Knob  as  shown 

below. [When you need to save the changes] 

 

If you don’t want to save it, press the Stirrer Knob when display shows 

“SAVE NO”. [When you don’t need to save the changes] 

 

Step 5: Terminate the configuration mode as described in Configuration 

Mode. 

 

 

•  When  the  heat  up  time  is  too  slow,  check  whether  the  maximum 

heating rate is set too low. 

•  Maximum  heating  rate  set  point  can  only  be  applied  in  top  plate’s 

control mode, not the external temperature probe mode. 

 

Changing the High 

Temp Limits (t-H) 

 

Note that the default setting of the high temperature limit [t-h] 350°C. If 

needed,  however,  you  can  set  your  own  temperature  limits.  If,  for 

example,  you  want  to  set  the  high  temperature  limit  at  200°C,  do  as 

follows: 

 

Step 1: Press the CONFIG Button.

 

 

Step  2:  select  the  temperature  limit  mode  shown  below  by  turning  the 

Stirrer  Knob.  Press  the  Stirrer  Knob  and  check  the  current 

high temperature limit: 

 

 

4-20

 

 

Section 4   

Control Panel

 

 

 

Содержание Fisherbrand Isotemp RT AVCD HPS Advanced

Страница 1: ...Operation Manual Isotemp RT AVCD HPS Advanced Hotplate Stirrer 9240 11 012 V01 12 01 17...

Страница 2: ...This manual cover the model is shown below NA Model EU Model Voltage Description 11676261 15326607 230V EU UK ANZ CN Isotemp RT AVCD HPS230V 11676262 N A 120V US Isotemp RT AVCD HPS120V...

Страница 3: ...fications appearance and dimensions described in this document without notice Fisher makes no representations or warranties with respect to this manual In no event shall Fisher be held liable for any...

Страница 4: ...hese alerts are classified into four types in this manual depending on the importance and the risk levels as described below Symbols Meaning Indicates a hazardous situation which if not avoided will r...

Страница 5: ...g the Temperature Probe 4 3 Stirrer Knob 4 5 Run Stop Button 4 5 Temperature Setting 4 6 Temperature Setting During Operation 4 7 Resetting Set Temp During Operation 4 8 Maximum Heating Rate Setting 4...

Страница 6: ...Escape Mode ESC 4 29 Section 5 Safety Device 5 1 Section 6 Maintenance 6 1 Cleaning Product 6 1 Relocation 6 2 Keeping Product 6 2 Section 7 Trouble Shooting 7 1 Error Messages 7 3 Section 8 Accessori...

Страница 7: ...compound Inflammable materials salt peroxides inorganic peroxide salt acids Check equipment for permissible environmental condition It can be the cause fire or trouble by electricity electronic and da...

Страница 8: ...ns near this equipment Do not touch the power outlet power socket or power cord with wet hands And make sure to connect the power cord directly and firmly to the power outlet and power socket Do not p...

Страница 9: ...with viscous media or at high speeds 30 2000rpm guaranteed The BLDC motor and the special magnet also provide big stirring power Real time stirring with quick operation or stop When the external temp...

Страница 10: ...eramic coated top plate is highly resistant to heat and corrosion In addition its white color is optimal for monitoring color changes of the media during operation Safety Hot Top Warning Indicator The...

Страница 11: ...bath during operation this bath has concavo convex bottom shown below so that it can sit stably on top of the top plate Figure 2 1 Heating Bath Convenience Conveniently set indicating temperature and...

Страница 12: ...of devices such as temperature sensors thermometers laboratory glassware etc optional User designation temperature range Your own temperature limit can be easily set and protect simple from controlli...

Страница 13: ...obe This is PT100 sensor to check media temperature optional 2 different temperature probe series Refer to Accessories list 2 Heat sink 3 Heater knob 4 Control panel 5 Stirrer knob 6 Clamp holder opti...

Страница 14: ...2 5 Section 2 Functional Description Construction Figure 2 3 Back Components 10 Power switch 11 Transparent Shield optional 12 Top plate 13 Threaded hole 14 Power socket 15 Temperature probe cap...

Страница 15: ...all the following parts and accessories are included in the package If not contact your dealer or Fisher Scientific immediately Item Figure Quantity Description Main body 1 Power cord 1 Operation Man...

Страница 16: ...are as follows 1 Before connecting the power cord make sure that the main power switch is turned off 2 Plug the power cord into the power socket at the back of your instrument as shown in the diagram...

Страница 17: ...tion STIRRER During stirrer operation PROBE During the Probe measurement temperature indication SET During temperature Timer Config set WARNING Overheating warning C Temperature indication display Hot...

Страница 18: ...ighest temperature set top plate and external temperature sensor t L Lowest temperature set top plate and external temperature sensor List Indicating the set value to turning stirrer knob Calibration...

Страница 19: ...switch the following display will appear in sequence Connecting the Temperature Probe When the external temperature probe connects with body the temperature display part shows Probe measuring tempera...

Страница 20: ...robe from main body the temperature display windows are showed follow message When you connect with temperature probe check first hotplate temperature When you connect with temperature probe and hotpl...

Страница 21: ...rpm you can control more correctly from 30 to 2000rpm Run Stop Button Even if the main power switch is turned on you have to press the RUN STOP Button to start or to finish the operation The RUN LED i...

Страница 22: ...in very hot for some time Never leave your instrument accessible to others while it is hot and never touch it unless you are absolutely sure You can check set temperature during heater operating Press...

Страница 23: ...Temperature Setting Step 3 Push the RUN STOP Button Heater LED will be on and heater will be activated If the external temperature probe is connected the temperature which probe detects will be displ...

Страница 24: ...see the set temperature If you press TEMP Button set temperature is shown on the right After 15 seconds the mode will be escaped automatically If you press the Heater Knob set temperature is shown on...

Страница 25: ...turning and press the Heater Knob After 15 seconds the mode will be escaped automatically Or push RUN STOP Button as shown below to escape the mode Step 1 Push RUN STOP Button during resetting the set...

Страница 26: ...s 100 but you want to set the heating rate limit at 50 do as follows Step 1 Press the CONFIG Button The default value is 100 Step 2 Change the limit to the desired value by turning the Stirrer Knob an...

Страница 27: ...to begin the timer setting and also to check the display showing the current timer setting Note that the hour frame is blinking first as shown below Step 2 If you want to change the timer setting sel...

Страница 28: ...firmation as shown below Operating confirmation Step 5 When the timer operation ends you will be alerted by audible signals as well as the following display Terminate the timer operation either by pre...

Страница 29: ...er mode is shown as below Step 1 Press RUN STOP Button Step 2 Turn the Stirrer Knob clockwise and press the Stirrer Knob Temperature Probe Control Mode Connect temperature probe to the unit The defaul...

Страница 30: ...sensor to completely immerse into the media more than 20mm deep If the sensor is in the air due to evaporation of media precise temperature control may fails Refer to Error Message Error code Cause Ef...

Страница 31: ...erminate the configuration mode by using one of the following methods Step 1 Press the RUN STOP Button Step 2 Turn the stirrer knob to select YES and push it to confirm the termination as shown below...

Страница 32: ...ning and pressing the Stirrer Knob Step 4 Save the selected temperature unit by using the Stirrer Knob as shown below When you need to save the changes If you don t want to save it press the Stirrer K...

Страница 33: ...erature is reached The timer mode selection procedures are as follows Step 1 Press the CONFIG Button Step 2 Select the timer mode shown below by turning and pressing the Stirrer Knob Step 3 If you wan...

Страница 34: ...nces at 0 1 C interval Offsetting of top plate is displayed as OFST and offsetting of the external temperature probe is displayed as OFSP The allowed range of the temperature offsetting OFST 10 50 C O...

Страница 35: ...Knob The allowed ranges of the temperature offsetting are from 10 to 50 C for OFST and from 3 to 3 C for OFSP OFST Top plate offsetting OFSP External temperature probe Step 4 Save the desired offsetti...

Страница 36: ...fore if you want to reduce the temperature fluctuation you need to limit the heating rate to a certain degree by limiting maximum heating rate Same function as Temperature Setting If for example the c...

Страница 37: ...ting rate is set too low Maximum heating rate set point can only be applied in top plate s control mode not the external temperature probe mode Changing the High Temp Limits t H Note that the default...

Страница 38: ...If you don t want to save it press the Stirrer Knob when display shows SAVE NO When you don t need to save the changes Step 5 Terminate the configuration mode as describe in Configuration Mode The all...

Страница 39: ...ep 2 Select the temperature limit mode shown below by turning the Stirrer Knob Press the Stirrer Knob and check the current low temperature limit Step 3 Change the limit to the desired value by turnin...

Страница 40: ...ngs as follows Step 1 Press the CONFIG Button Step 2 Select the configuration checking mode LiSt shown below by turning the Stirrer Knob Step 3 Press the Stirrer Knob and check the current settings by...

Страница 41: ...circumstances into consideration If you select User Mode Coefficient the unit will be operated with the auto tuning value The auto tuning calibration procedures are as follows Step 1 Press the CONFIG...

Страница 42: ...utomatically with an alarm Coefficient COEF Five user selectable temperature control modes are provided for your convenience Step 1 Press the CONFIG Button Step 2 Select temperature control mode selec...

Страница 43: ...n range of temperature overshoot and undershoot Point Mode Allowing linear heating up to the target temperature using the maximum heating rate and switching on and off the heater based on the target t...

Страница 44: ...ollows Step 1 Press the CONFIG Button Step 2 Select the default setting dEFA shown below by turning the Stirrer Knob Step 3 Press the Stirrer Knob and confirm the change to the default settings by usi...

Страница 45: ...it of maximum stirring speed The range of speed adjustable is from 1 to 1000rpm Step 1 Press the CONFIG Button Step 2 Select the Motor Minimum Mmin shown below by turning the Stirrer Knob Step 3 Chang...

Страница 46: ...hen display shows SAVE NO When you don t need to save the changes Step 5 Terminate configuration mode as described in Configuration Mode Escape ESC Terminate the configuration mode Step 1 Press the CO...

Страница 47: ...and visible signals will be activated 4 In such cases turn off the power switch and disconnect the power cord first Then allow your unit to cool down completely before operating it again 5 Hot Top War...

Страница 48: ...f any overheat prevention activates or warning sound is alarming turn off the power switch and disconnects the power cord first Then allow your unit to cool down completely before operating it again 5...

Страница 49: ...f all parts or accessories Cleaning Product Never immerse this unit in water or any other liquid Do not allow any liquid or wet material to get inside the unit when cleaning Do not reconnect this unit...

Страница 50: ...easier if spills are attended to promptly Relocation If you need to move the instrument to another place 1 Disconnect the power cord from the power outlet 2 Pack the instrument and its accessories in...

Страница 51: ...or plugs Replace the damaged part with a proper one Internal circuit failure Contact Fisher Scientific for service Repetitive tripping of the circuit breaker Electrical overload Disconnect all the app...

Страница 52: ...increase the heating level Too much media Reduce the media volume Internal circuit failure Contact Fisher Scientific for service Stir bar decoupling Too much media Reduce the media volume or increase...

Страница 53: ...t rest enough Error Code Cause Effect ERRO NO 1 Related to Heater system If Top plate is overheated Heater off ERRO NO 2 If sensor detects a sudden temperature changes Heater off ERRO NO 3 If PCB is o...

Страница 54: ...to 400 11676269 15366607 FS Heating Bath 11676270 15376607 FS Transparent Shield PC 11676271 15386607 FS Support Rod 12 support 400mm M10 11676276 15346617 FS Clamp Holder PP body 12mm 11676277 153566...

Страница 55: ...8 1 Section 8 Accessories Transparent Shield Heating Bath Assembly When necessary the transparent shield and or the heating bath can be easily assembled as shown below...

Страница 56: ...ype BLDC Magnetic Stir Bar Max x L mm inch 8 x 40 0 31 x 1 57 Probe Type PT100 B class Accuracy C at 0 C 0 3 Timer 2 type 1min to 99 hr 59 min Safety Hot top warning C F Warning lamp 50 122 Overheat P...

Страница 57: ...or instrument disposal may lead to actions against you and your organization 1 First check with your laboratory or organization to ensure that you are following all the policies and procedures for dis...

Страница 58: ...ions and according to these operation instructions and maintenance this product is warranted for 24 months against defective materials or workmanship The 24 month warranty period begins from the deliv...

Страница 59: ...10 1...

Страница 60: ......

Страница 61: ...59 fishersci no Denmark 45 70 27 99 20 fishersci dk Portugal 351 21 425 33 50 pt fishersci com Germany 49 0 180 5258221 de fishersci com Spain 34 002 239 303 es fishersci com Ireland 959 0 1 885 5854...

Страница 62: ...Find out more at fishersci com 2017 Thermo Fisher Scientific Inc All rights reserved Trademarks used are owned as indicated at fishersci com trademarks...

Отзывы: