Changing the Low Temp
Limits (t-L)
Step 5: Terminate the configuration mode as described in Configuration
Mode.
• The allowed range of the temperature limit setting is 0°C ~ 350°C.
• However, low temperature limit [t-L] set point should not be more than
high temperature limit [t-H] set point.
Check Configuration
Settings (LISt)
You can check the current configuration settings as follows:
Step 1: Press the CONFIG Button.
Step 2: Select the configuration checking mode [LiSt] shown below by
turning the Stirrer Knob.
Step 3: Press the Stirrer Knob and check the current settings by turning
the Stirrer Knob.
The default configuration settings are as follows:
MODE
Default Setting
MODE
Default Setting
Unit
Cels
Time
Ti1
IFST/OFSP
0.0
Limt
100.0
t-H
350.0
t-L
0.0
Coef
Opti
Mmin
1
Step 4: After checking the settings, terminate the checking mode.
[Either by pressing RUN/STOP Button]:
[Or by pressing the Stirrer Knob]:
4-23
Section 4
Control Panel