5
1. Before Started
1.1 Metrological Legislation
Because of metrological legislation, installation and some metrological
parameter settings / changings are limited to be done by authorized
personnel only. Do not attempt to change any of the built-in parameters.
Contact your dealer for installation and technical assistance.
1.2 Seal & Serial Number
This scale is legal for trade only when it is sealed (and/or stamped) and
bearing a serial number. Do not attempt to break the seal (or stamp) or
serial number affixed to scale. Contact your dealer for more information and
after sales service.
1.3 Warm Up Time
a.
Allow warm up period of not less than 60 seconds before calibration.
The higher the setup resolution of the scale, the longer the warm up
period is required. In most cases, 120 seconds is a safe warm up
period for all applications.
b.
This warm up period is needed to energy all components to reach a
stable status.
c.
The internal count value is deemed stable when the internal AD
count varies less than 3 counts within 2 seconds.
d.
To read the internal AD count value, enter internal function F1. The
internal AD count value of a not yet fully energized PCB will go up
continuously.
1.4 In Case When in Doubt
Always contact your dealer for more information, after sales service and
questions when in doubt.