When  F13  is  set  to  ON,  scale  will  permit  multiple  tare  operation  provided 

that both of the below requirements are met: -   

 

a.  The tare operation does not permit a reduction of the value of the tare;   

b.  The  tare  effect  can  only  be  cancelled  when  there  is  no  load  on  the 

platform. 

 

6.4.4 Preset Tare (F63)

10

 

A pre-determined tare  weight can be  entered  via keyboard. To enable this 

function, set F63 to ON, then set also F12 to OFF. 

 

a. 

During  normal  operation,  press 

[Func] 

until

  PtArE 

appears,  then 

press

  [Print]

  to  enter.  Input  the  pre-determined  tare  weight  through 

[M+]

[Unit]

 and 

[Func]

 keys and press

 [Print]

.   

 

To  cancel  preset  tare  effect,  remove  all  loads  from  platform  then  press 

[Tare]

 

Note: - 

 

1. 

The pre-determined tare weight entered will be rounded to the nearest 

division  of  the  instrument.  This  does  not  affect  the  accuracy  of  the 

subsequent weighing and operation. 

2. 

Manual tare is possible when preset tare is in function. 

3. 

Preset Tare is also governed by Repeated Tare (F13) 

 

6.5 Memory Accumulation Function 

6.5.1 To Accumulate a Transaction to Memory

11

 

12

 

13

 

 

a.  Press 

[M+]

14

  to  save  and  accumulate  data  of  current  transaction  to 

memory, 

                                                      

10

    This is not a legal for trade function. 

11

    Memory Accumulation Function support weighing function only. 

12

    Weight less than 20d (or 20d

for dual range) will not be accumulated to memory. 

13

    All data stored will be erased when scale is powered off. 

14

    or press 

[PRINT]

 if F17 is = ON 

Содержание DM-180 SERIES

Страница 1: ...1 DM 180 SERIES DIGITAL WEIGHING SCALE OPERATION MANUAL PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THIS SCALE V100 Feb 2013 Specifications subject to change without prior notice ...

Страница 2: ...2 ...

Страница 3: ...er And Select Internal Function 13 5 3 Key Function During Internal Function Mode 14 5 4 Internal Function Table 14 6 Instruction for Use 19 6 1 Power On 19 6 2 Start Weighing 19 6 3 About Weigh Unit Conversion 19 6 4 Tare off the Weight of a Container 20 6 5 Memory Accumulation Function 21 6 6 Piece Count Function 22 6 7 Sampling Process 23 6 8 Shift among Quantity Average Piece Weight and Weight...

Страница 4: ... Label File Names 29 9 2 Label Programming 30 10 Battery Power Recharging 33 10 1 Symbols And Remaining Power 33 10 2 Battery Operation Time 33 10 3 Recharge Battery 33 11 Error Codes 34 12 Daily Care Maintenance 35 Appendix A Bi Directional Communication Commands 36 ...

Страница 5: ...ormation and after sales service 1 3 Warm Up Time a Allow warm up period of not less than 60 seconds before calibration The higher the setup resolution of the scale the longer the warm up period is required In most cases 120 seconds is a safe warm up period for all applications b This warm up period is needed to energy all components to reach a stable status c The internal count value is deemed st...

Страница 6: ...its kg g lb Weight Unit Conversation Supports Single Range Models only Max Tare Range Max Subtractive Tare Platter Cover 220 x 290mm all Stainless Steel Dual Power Source Built in Rechargeable Battery 6V 4AH External Power Adaptor DC 12V 1A with Universal Plugs Operation Environment 10 40oC Non condensed R H 85 Specifications subject to change prior to notice ...

Страница 7: ...eys Display Connections 0 Spare Spare key no assignment 1 Spare Spare key no assignment 2 Zero Key Press this key to set weight displayed to zero when an empty scale has drifted away from a true zero reading ...

Страница 8: ...accumulate current weight to memory manually 9 Battery Power Level Indicator Visible to show Scale is being powered by the built in rechargeable battery Remaining battery level 10 W1 Indicator3 When under dual weighing range mode4 Visible when scale is in the first weighing range W1 1 When F11 ON 2 This key is also used to accumulate the current weight value to memory when internal function F17 is...

Страница 9: ...wn is net value 15 Stable Indicator Visible when weight reading is stable 16 Zero Indicator Visible when instrument is at true zero weight status 17 M Indicator Visible when memory contains of accumulated data 18 MR Indicator Visible when the total accumulated weight value is displayed 19 Weight Units and Functions kg kilogram PCS Pieces Piece Count Mode in function kg PCS and g PCS Weight per pie...

Страница 10: ...or Charging completed BOTTOM VIEW A On Off Key Press this key to turn this scale on or off B DC Jack Input External power adaptor is plugged in here Output requirements of the power adaptor DC9 12V 800mA Polarity Any kind C RS232 Comport DB9 RS232 communication comport ...

Страница 11: ...vibration or extremes of temperature before or after installation 4 1 Built In Rechargeable Battery This scale is equipped with a built in rechargeable battery Before first time use recharge it for at least 8 hours to ensure the best battery performance 4 2 Power Adaptor One universal voltage power adaptor with 4 power plugs are supplied with scale Select the suitable power plug and slide it into ...

Страница 12: ...eferred Operation Parameters Set all preferred operation parameters according to 5 4 INTERNAL FUNCTION TABLE Note 1 F1 F29 are accessible without restriction 2 F60 F66 are restricted functions which may request a password or hardware key to access 3 F80 F99 are restricted functions which may request a password or hardware key to access These functions are usually for dealer and authorized personne...

Страница 13: ... function Follow the below steps to enter and select desired parameter of an internal function For F60 F99 If Setup Calibration Control Jumper is set to enable position B C instrument will enter internal function automatically after power on Should password mode be used to bypass this jumper follow the below steps to enter and select desired parameter of an internal function a Power scale off and ...

Страница 14: ...countdown or set F1 value being shown to zero and to display the net span gain by applying additional load applied Func Goto next page Move cursor to one place right Unit Goto previous page Move cursor to one place left Print Enter save and return MR Clear M Increase numeric number by 1 5 4 Internal Function Table Refer to the below table for internal function number parameter and setting notes ...

Страница 15: ...15 ...

Страница 16: ...16 ...

Страница 17: ...17 ...

Страница 18: ...18 ...

Страница 19: ...ds from platform after weighing It will prolong the life of the weight sensor 6 3 About Weigh Unit Conversion The default weight unit is kg Depends on the internal settings scale supports also g and lb 6 3 1 Conversion between Metric Weight Units kg and g When 3 or 4 decimal places 0 000 or 0 0000 is selected in F80 reading in g is possible during normal operation by pressing Unit disregarding to ...

Страница 20: ...r appears to indicator tare is in effect and weight displayed display is net weight f To cancel tare effect remove all loads from platform and press Tare g Net Indicator disappears Gross Indicator appears to indicator tare effect has been removed and weight displayed display is gross weight 6 4 2 Auto Tare8 F12 If this function is enabled scale will assume the first stable weight applied is a cont...

Страница 21: ...l preset tare effect remove all loads from platform then press Tare Note 1 The pre determined tare weight entered will be rounded to the nearest division of the instrument This does not affect the accuracy of the subsequent weighing and operation 2 Manual tare is possible when preset tare is in function 3 Preset Tare is also governed by Repeated Tare F13 6 5 Memory Accumulation Function 6 5 1 To A...

Страница 22: ...f transactions accumulated follow by the total accumulated weight stored in memory b At this point Press Zero to quit or Press MR again to clear memory then press Zero to quit to operation mode After that both MR and M Indicator disappear to indicate no data is stored in memory 6 6 Piece Count Function16 This instrument is equipped piece count function Follow the below steps to enter Piece Count F...

Страница 23: ... and 6 7 6 8 Shift among Quantity Average Piece Weight and Weight Info a Press Unit to shift among quantity average piece weight and weight info b Quantity Display format numeric numbers PCS e g1000 PCS c Average piece weight display format numeric numbers weight unit slash PCS e g 499 960g PCS d Weight display format when Piece Count Function is in effect numeric numbers weight unit PCS e g 500 k...

Страница 24: ...ransmitted in ASCII code Data format is listed on below table DATA BIT DESCRIPTION 1 2 MOTION STATUS US UNSTABLE ST STABLE 3 COMMA SEPARATION 4 5 NET GROSS NT NET WEIGHT GS GROSS WEIGHT 6 SIGN Sign of weight reading Positive space Negative minus 7 13 WEIGHT VALUE 7 digits weight value including location of decimal point If there is no decimal point then the first character space 14 COMMA SEPARATIO...

Страница 25: ...mat 8 1 1 1 Weighing function 7 lines will be transmitted as below 1 Time of print 2 Date of printing 3 Transaction sequent number if this transaction is accumulated to memory 4 Net weight 5 Tare Weight 6 Gross Weight 7 Total accumulated net weight if accumulation function is in effect Sample 1 TIME 15 21 00 DATE 14 04 2009 NO 1 First transaction added to memory NET 500 0kg TARE 0 0kg GROSS 500 0k...

Страница 26: ...09 NO 3 Third transaction added to memory NET 500 0kg TARE 200 0kg GROSS 700 0kg TOTAL 1200 0kg Total accumulated net weight 8 1 1 2 Piece count function 5 lines will be transmitted as below 1 Time of print 2 Date of printing 3 Net weight 4 Unit weight average piece weight 5 Count quantity in terms of number of pieces Sample 1 TIME 15 30 44 DATE 14 04 2009 NET 300 0kg UNIT W 599 949 g COUNT 500PCS...

Страница 27: ...d the last variant or command see 8 2 2 for details stored d Press M to confirm or select other variant or command by press FUNC or UNIT Then press M to confirm and save e Scale displays Line 2 and the last variant or command stored f Repeat steps d and e for other lines g In case number of lines to be printed is less than 15 lines To finish editing select command End then pres M to confirm h Scal...

Страница 28: ...xt line dAtE Date of printing tiME Time of print nEt Net weight tArE Tare weight GroSS Gross weight Unit Average piece weight cOuNT Number of piece ACC Total accumulated weight when accumulation function is in effect SiGn Signature 8 2 3 Edit Sample for Custom Print Output Format PRINT CONTENT Line No Select TIME 17 39 05 1 tiME DATE 14 04 2009 2 dAtE NET 200 0kg 3 nEt TARE 0 0kg 4 tArE GROSS 200 ...

Страница 29: ...wrong data of non current working mode 9 1 Label Format Groups Label File Names 2 label format groups are available these are for 2 label format group 1 and For 2 label format group 1 9 1 1 For 1 Label Format Group 1 For 1 format group 1 is for current transaction data printing during normal working status In order to trigger the right label to be printed label files stored in printer for this for...

Страница 30: ... label file BB1 dlb stored in printer For 2 2 Select this to print label file BB2 dlb stored in printer For 2 3 Select this to print label file BB3 dlb stored in printer For 2 4 Select this to print label file BB4 dlb stored in printer For 2 5 Select this to print label file BB5 dlb stored in printer 9 2 Label Programming Prompt commands information description working mode and suggested length on...

Страница 31: ...d transaction Weighing 7 N Total accumulated weight Weighing 9 O Net weight All 10 P Tare weight All 10 Q Gross weight All 10 U Number of piece Piece Count 10 V Average piece weight Piece Count 9 9 2 2 Label Programming Sample 9 2 2 1 Sample Label of Current Transaction For 1 28 Commands must be in capital letter 29 All means the information is good for all working modes ...

Страница 32: ...32 9 2 2 2 Sample Label of Totalized Data For 2 ...

Страница 33: ...echargeable battery is at low voltage status It is recommended to recharge as soon as possible To protect the built in rechargeable battery scale will be powered off automatically when battery is at extremely low level If this is the case do not attempt to power scale on Recharge scale immediately Fail to do so may cause unrecoverable damages to the built in rechargeable battery Battery charging s...

Страница 34: ...r on Err 5 No load cell signal detected Err 6 Tare operation error Err 7 Reserved no definition Err 8 Reserved no definition Err 9 Reserved no definition Err 10 Reserved no definition Err 11 Reserved no definition Err 12 Reserved no definition oL Overload Gross weight is more than Max plus 9d UndEr Negative Weight values exceeds display range Negative Tare value exceeds display range ...

Страница 35: ... material acetone volatile solvent thinner or alcohol for cleaning Verify the accuracy of scale periodically Re calibrate if necessary In some countries calibration requires authorized qualified agent Contact your dealer for more information Store scale in a dry and clean place Recharge battery before and every 2 months during long time storage ...

Страница 36: ...h any standard communication programs like Hyper Terminal which comes with most of the Windows operating system in computer A Direct Control Commands31 F18 Mode 1 or 2 Equivalent direct control commands for each of the key on panel from computer are shown below 31 Direct Control Commands to be transmitted to scale are not case sensitive ...

Отзывы: