10 - MAINTENANCE
10.2 Disinfecting the device
DISINFECTING THE NON-MOTORISED DEVICE
(VENICE BASIC AND VENICE PLUS)
1. Remove restraints and any other accessory from the
product.
2. Disinfect all equipment (including accessories such as
cushions, restraints, etc.) with a cloth or sponge dampened
with disinfectant solution, let it sit for 10-20 minutes; use
brushes to rub and clean the least accessible parts;
3. Use a cloth to rinse and dry the treated surfaces, make
sure that all the device's components are completely dry
before using them;
4. After cleaning, be sure to refit on the device the restraints
and/or any other accessory that was removed.
10.1 Routine and preventive maintenance
The chair is a medical device requiring regular maintenance,
hence you are recommended to comply with the following
schedule:
1. ROUTINE maintenance: Ferno recommends carrying
out routine maintenance operations once a year. Ferno
engineers will issue a conformity certificate of the medical
device after maintenance.
2. PREVENTIVE maintenance: preventive maintenance
operations are to be carried out by the user, as specified in
the table.
If the chair undergoes particularly heavy use, maintenance
should preferably be carried out on a more frequent basis. If
there are signs of wear and damage, discontinue use of the
chair and immediately contact Ferno's Technical Support
Service.
Ferno directly carries out repair maintenance operations,
without the need of dealers, mechanics or external service
centres.
You are thus invited to be wary of those who present
themselves as Ferno's authorised technician. We suggest
you to directly contact our SOS service using the reference
contacts specified on page 2.
Important
Disinfectants and detergents containing phenols (if
above the threshold of 2500 ppm), chlorine-based
(bleach) or iodine-based solutions can cause damage.
Do not use products containing these chemicals.
Important
When not in use, store the device in an indoor
environment that is dry and protected from direct
sunlight.
Important
DO NOT WASH WITH HIGH PRESSURE
WATER
Do not use high pressure water to clean the
device.
Important
After disinfection / cleaning, ensure that the device
and its components are completely dry before use.
Disinfection
Cleaning
Inspection
Lubrication
- At least every 6 months
Voltage adjustment
- At least every 6 months
At every use
Every month
When needed
Minimum preventive
maintenance intervals