Basic Training Workbook
Version 1.5 - August 2007
12
Chapter 1: Getting Started
C
AUTION
:
The probe
must
be well-seated in the hole when digitizing all calibration points.
Even one or two poorly digitized points can significantly affect the optimization process,
which may reduce the accuracy of the FARO Gage.
The calibration points are then calculated and the probe calibration status is updated. If the
probe passes, then the current date and time is added to the probe information.
12 You may view the details for all of the calibrations of a probe by selecting the VIEW LOG
button.
•
If you calibrate the probe correctly, the calibration status changes to PASS.
•
If the probe calibration fails, select the RETRY button and repeat the calibration until
the probe passes.
You are now ready to measure with the FARO Gage.
11 Digitize points in the hole and sweep to
position #3.