FTP9000 series power supply allows for output voltage ramp-up and ramp-down according to the set slew rate.

Operation:

1. Press

to enter test modes selection interface;

2. Use “Rotary”

knob to select “RAMP”, press

, the screen displays as below:

80V/510A/15000W

RAMP

1/3

0.0000

V

0.0000

A

0.000

w

Voltage A

0.000 V

Voltage B

0.000 V

Voltage C

0.000 V

Off

80V/510A/15000W

RAMP

2/3

0.0000

V

0.0000

A

0.000

w

V

slew rate

- AB

0.000 V/s

V

slew rate

- BC

0.000 V/s

I_Set

0.000 A

Off

80V/510A/15000W

RAMP

3/3

0.0000

V

0.0000

A

0.000

w

P_Set

0.000 W

Off

Figure 3-17 RAMP Function Parameter Set

3. Set the corresponding parameters, you can refer to “Table 3-6 RAMP Function Parameter Explanation” ;

4. Press

to turn on voltage ramp output.

Table 3-6 RAMP Function Parameter Explanation

Parameter

Explanation

Содержание FTP9000 Series

Страница 1: ...Programmable DC Power Supply User s Manual FTP9000 Series Copyright Faithtech Version V1 03 2021 06...

Страница 2: ......

Страница 3: ...date of purchase warranty within the one year warranty period Faithtech will provide free repair for all faults in the normal usage and maintenance of the product The user need prepay one way freight...

Страница 4: ...n Make sure the output wire meets the specified standards and a protective cover must be installed to prevent electric shock Do not open the instrument case Operators are not allowed to open the shell...

Страница 5: ...tly observed could result in injury to or death of personnel or long term heath hazards The CAUTION sign highlights an essential operating or maintenance procedure practice condition statement etc whi...

Страница 6: ...18 2 6 3 Power ON Self Check 18 2 7 Connection 19 2 7 1 Input Connection 19 2 7 2 Input Connection for Multiple Power Supplies 19 2 7 3 Output Connection 20 2 7 4 Remote Sensing Connection 20 2 7 5 PR...

Страница 7: ...mp down RAMP 38 3 15 Master slave Parallel 40 3 15 1 Master slave parallel connection 40 3 15 2 Master slave parallel setting 41 3 15 3 Master slave parallel function operations 42 4 COMMUNICATION INT...

Страница 8: ...fit from the standard USB virtual SERIAL and LAN interfaces supporting both SCPI commands and ModBus RTU protocol Free application software programming manual and a complete set of development DLLs ar...

Страница 9: ...____________________________________________________________________________ 7 1 3 Dimension diagram 1 3 1 Model of 5kW 10kW 15kW...

Страница 10: ...FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ 8 1 3 2 Model of 20kW 25kW 30kW...

Страница 11: ...l GPIB CAN RS485 Protocol SCPI MODBUS CAN Open protocols Input characteristics Input voltage 340VAC 460VAC 47Hz 63Hz Power factor 0 99 Typical Efficiency 93 Typical Operation environment Working temp...

Страница 12: ...300A Power 0 60kW Model FTP9900 80 3060 FTP9900 200 1260 FTP9900 400 720 FTP9900 500 540 FTP9900 800 450 Voltage 0 80V 0 200V 0 400V 0 500V 0 800V Current 0 3060 A 0 1260A 0 720A 0 540A 0 450A Power...

Страница 13: ...200 40 FTP9150 1500 30 FTP9150 2250 25 Voltage 0 1000V 0 1200V 0 1500V 0 2250V Current 0 40A 0 40A 0 30A 0 25A Power 0 15kW Model FTP9200 1000 60 FTP9200 1500 50 Voltage 0 1000V 0 1500V Current 0 60A...

Страница 14: ...F S Current accuracy 0 5 F S Line regulation Voltage 0 02 F S Current 0 05 F S Power 0 05 F S Load regulation Voltage 0 05 F S Current 0 15 F S Power 0 75 F S Voltage measurement Resolution 16bits Ac...

Страница 15: ...utput protection 2PIN 5 08mm terminal 1 PCS Voltage remote sense connection terminal 2PIN 2 54mm terminal 1 PCS Master Slave parallel connection terminal USB communication cable 1 PCS For connecting P...

Страница 16: ...l Power switch TFT color screen Keyboard Rotary knob 2 3 Keyboard Figure 2 2 FTP9000 keyboard Table 2 2 Key Function Descriptions Button Key Name Description Voltage Setting Key Switch to output setti...

Страница 17: ...lear Key Clear the occurred protection and warning messages Main Menu Key Enter the main menu Compound Function Key Work with other key together Lock or unlock the keyboard In locked state only and ar...

Страница 18: ...ters display area Displays readback voltage current power measurement result etc However the actual displayed information varies for example in V I P static output mode the area displays output voltag...

Страница 19: ...negative output terminal Do not reverse the connections 3 Output The output terminals of the DC Power Supply 4 GPIB or CAN RS485 The remote controller uses the GPIB connector to connect to a PC for r...

Страница 20: ...pply need to be connected to L1 L2 L3 and grounding terminal Please refer to Table 6 1 Recommended Wire Diameters for Cables for the selection of input wire diameters to avoid potential safety hazards...

Страница 21: ...to the earth Under no circumstances shall this DC Power Supply be operated without an adequate ground connection 2 Installation of the power cord must be done by a professional and compliant with loc...

Страница 22: ...dequate so as to carry the maximum applied current 2 7 4 Remote Sensing Connection FTP9000 series power supplies are able to compensate for the voltage drop caused by output connecting cables Use two...

Страница 23: ...ol 4 DGND Digital ground 5 DGND Digital ground 6 AGND Analog ground 7 VMON Output Output Voltage monitoring port monitoring output range 0 5VDC corresponds to 0 100 F S rated voltage of the power supp...

Страница 24: ...e by default In this mode the user operates the power supply through the front panel keyboard The LCD screen provides users with display functions such as parameter viewing measurement display and sta...

Страница 25: ...Set communication error time out range 0 0 60 0s set 0 to turn off this protection Detect RV Turn on off reverse voltage protection default value Closed Limit V_MAX Upper limit of static output volta...

Страница 26: ...0 254 default 160 GPIB set GPIB address range 0 30 default 5 Protocol SCPI MODBUS default MODBUS Factory Reset to factory setting Reboot the machine to take effect Others Language Simplified Chinese...

Страница 27: ...Press power supply enters static output mode the cursor for voltage set blinks also you can rotate the knob to highlight the voltage setting and press to edit voltage 2 Use to set the value and press...

Страница 28: ...e in which the output power can be changed by setting the output power value in this situation CP is displayed in status prompt area 3 3 2 Limit Voltage Current Power Set FTP9000 series power supply a...

Страница 29: ...as below Set System Edit About Samp Rate 10Hz Prio SLT Volt Vslew UP 5000 0 V s Vslew DN 5000 0 V s Islew UP 2000 0 V s Islew DN 2000 0 V s Figure 3 3 Voltage Current Slew Rate Set 3 Select parameter...

Страница 30: ...priority while Curr stands for current priority 3 6 Output Control The power supply s output is turned off when it is switched on Users control its output via front panel key key is lighted suggests...

Страница 31: ...XT S Trigger Figure 3 8 External Programming Set 3 Select APG_V APG_I APG_P for performing voltage programming current programming or power programming turn on off operations respectively the status p...

Страница 32: ...ree behaviors Trigger Default setting Used as trigger signal The port detects a valid low level pulse that is a trigger signal is received Toggle Used as power supply output ON OFF control signal The...

Страница 33: ...11 Output Voltage Current And Monitoring Output 3 7 4 Output Mode Status Monitoring The FTP9000 series power supply has an output mode signal port and a fault indication signal port The port outputs a...

Страница 34: ...ervoltage protection hardware overvoltage OV and software overvoltage OVP Hardware overvoltage protection OV It is triggered once the output voltage exceeds 105 rated output voltage and OV is prompt o...

Страница 35: ...g time can be set as in Figure 3 12 Protection Set range 0 0 60 0 seconds Overtemperature protection It is triggered once the temperature of inner power components exceeds allowed temperature range OT...

Страница 36: ...position 2 voltage 24V current 20A power 2000W maximum voltage and current slew rate Shortcut Function 1 Press 2 Select Set System and press then select Others and press again 3 Set Shortcut as ON so...

Страница 37: ...to run cyclically for a specified times set to 0 means infinite loop Link can add other sequence file into test set to 0 means no linked sequence file When the sequence test is turned on the highest...

Страница 38: ...lly 6 Set Link press to validate it the editing cursor will move to SEQ Step automatically 7 Set SEQ Step press to validate it the editing cursor will move to V_Set automatically 8 Set V_Set press to...

Страница 39: ...e current sequence step Cycle shows how many cycles the sequence have run 3 13 Internal Resistance Simulation RESI Since a battery has internal resistance its output voltage will decrease along with t...

Страница 40: ...RESI 2 2 0 0000 V 0 0000 A 0 000 Resistance 0 000 Off Figure 3 16 Output Test For Battery Simulating 3 Set the corresponding test parameters 4 Press to turn on battery simulating output Table 3 5 RESI...

Страница 41: ...0W RAMP 1 3 0 0000 V 0 0000 A 0 000 w Voltage A 0 000 V Voltage B 0 000 V Voltage C 0 000 V Off 80V 510A 15000W RAMP 2 3 0 0000 V 0 0000 A 0 000 w Vslew rate AB 0 000 V s Vslew rate BC 0 000 V s I_Set...

Страница 42: ...rent P_Set Set output power range 0 Max output power V Voltage A Voltage C Voltage B t Figure 3 18 RAMP Function Parameter Set 3 15 Master slave Parallel 3 15 1 Master slave parallel connection FTP900...

Страница 43: ...unit if you want to compensate voltage line loss 4 Does not support master slave series operation 3 15 2 Master slave parallel setting Slave setting 1 Press to enter main menu interface 2 Select Set P...

Страница 44: ...rameter setting The master power supply specifications will be updated 3 15 3 Master slave parallel function operations Simply operate the master unit to control the whole group if you have configured...

Страница 45: ...the fault slave unit code requires user to press to clear the fault protection 1 Set slave unit first so as to avoid master unit communication error 2 Multiple power supplies have been set as master u...

Страница 46: ...rst please set IP as 192 168 1 123 subnet mask as 255 255 255 0 then set the IP address of the connecting PC to be in the same segment with the power supply IP Different IP segments will result in com...

Страница 47: ...2 Application Setting 3 Select required speed in Samp Rate there are 5 choices select able Default is 10Hz 4 3 Communication Interfaces 4 3 1 USB Serial FTP9000 provides a USB virtual serial port com...

Страница 48: ...address of the PC and the power supply are in the same network segment and that there is no duplicate IP address with the power supply in this network segment otherwise it will not be able to connect...

Страница 49: ...AN interface only supports CANopen protocol 4 4 1 SCPI protocol For specific content of SCPI protocol please refer to FTP9000 Programming Manual SCPI Edition 4 4 2 Modbus protocol For specific content...

Страница 50: ...ate Overtemperature Protection OTP1 OTP2 1 The ambient temperature is too high 2 Bad ventilation 3 Fan damaged 1 Operate the instrument within an ambient temperature of 0 40 C 2 Ensure the instrument...

Страница 51: ...ions AWG mm2 60 75 85 90 Wire type RUW T UF Wire type RHW RH Wire type V MI Wire type TA TBS SA AV Rated Current A 14 2 08 20 20 20 20 12 3 31 25 25 30 30 10 5 26 30 35 40 40 8 8 36 40 50 55 55 6 13 3...

Отзывы: