Adjustment
Page
26
/
47
Main Menu / Configuration // Output /// Variable
The variable to which to the 4 to 20 mA signal refers is determined here.
Main menu / Configuration // Output /// 4 mA point
Setting the 4 mA point of the probe (in units of length)
Main menu / Configuration
// Output /// 20 mA point
Setting the 20 mA point of the probe (in units of length)
Main Menu / Configuration // Output /// Alarm current
Current that is output in the event of an error: 3.6 mA / 21.5 mA.
Main Menu / Information // Device
Device type and device number
Main Menu / Information // Version
Firmware and hardware version of the probe
Main Menu / Information // Properties
Probe length, number of floats, number of temperature sensors.
Main Menu / Information // Communication
Shows which communication option the probe supports
Main Menu / Information // Calibration
Date of the last calibration of the device
Main Menu / Diagnosis // Status
Current status of the device
Main menu / Diagnosis // Min./Max.
Shows the minimum and maximum value that the probe has recorded since restart:
Level 1, Level 2, Temperature.