F.W. BELL 8000 Series Gauss/Tesla Meter Instruction Manual
Section 7 – Waveform Display
7-3
Setting the Time
Base (cont’d)
Selecting the appropriate AC bandwidth filter for the signal of interest will
allow the user to display the desired time resolution for the signal being
examined. Table 7-1 shows the time base settings for each of the AC
Frequency Filters.
For convenience the AC Freq Setup menu is duplicated in the Waveform
Setting menu. Refer to the AC Mode Operation section in this manual for
more description of the AC Frequency Filter Selections.
The selected time base setting is indicated in the lower left corner of the
display as seen in Figure 7-1.
Table 7-1 Time Base Selections for AC Filter Settings
High Freq
(20Hz -50kHz)
Med Freq
(20Hz -10kHz)
Low Freq
(20Hz -500Hz)
X Low Freq
(1Hz -500Hz)
Full BW &
No Correction
8
µ
s 16
µ
s 320
µ
s 320
µ
s 8
µ
s
24
µ
s 48
µ
s 640
µ
s 640
µ
s 24
µ
s
64
µ
s 128
µ
s
1.28ms 1.28ms 64
µ
s
400
µ
s 320
µ
s
1.92ms 3.20ms 400
µ
s
1.60ms 1.60ms 3.20ms 6.40ms 2.00ms
6.40ms 6.40ms 6.4ms 128ms 16.0ms
Setting the
Vertical Scale
From the Waveform Settings menu, accessing the Scale menu brings up
a row of hot keys that set the units per division measurement for the
vertical value of the waveform display.
The vertical scale selection range is associated with each channel’s input
range, such that one of six values is available to scale the display within
that channel’s selected range.
For example, with the range set to 30kG six selections ranging from
0.50kG to 20.00kG per division are available. For a range selection of
3mT there are six selections ranging from 0.03mT to 2.00mT per division.
Selecting the appropriate measurement range for the signal of interest
will allow the user to display the desired vertical scale for the signal being
examined. Refer to section 6-3 of this manual for more information
regarding measurement range selection.
The present measurement range and waveform scale per division
settings are indicated on the top of the waveform display as seen in
Figure 7-1.
Note: In waveform mode the range buttons for each channel are active
and can be used to change the range without using the menu system.