F.W. BELL 8000 Series Gauss/Tesla Meter Instruction Manual
Appendix C – Warranty
C-1
APPENDIX C Error Codes
Meter Specific Errors
Error
Code
Display Text
SCPI Text
0
No Error
No Error
1
The calibration data stored in the probe is not valid. Default values
will be used.
Invalid probe cal data
2
The calibration data stored in the meter is not valid.
Default values will be used.
Invalid meter cal data
3
The analog card is not present or not functioning.
Analog card not present
4
The temperature correction data stored on the probe is not valid.
Probe temp comp will not be used.
Invalid probe temp comp data
5
The 100mA AC and DC probe excitation currents are not within
tolerance.
100mA AC/DC Ic out of range
6
Error reading probe EEPROM.
Default values will be used
Error reading probe EEPROM.
7
This is a 7000 series probe. It will function but the readings will not
be as accurate.
7000 series probe attached
8
This is an unknown probe type. It cannot be used with this meter.
Unknown probe type
9
This probe has not been programmed. Default values will be used.
Missing probe cal data
10
This probe has not been programmed with temperature
compensation data. Probe temp comp will not be used.
Missing probe temp comp data
11
The requested range cannot be used while relative offset is greater
than its full scale value.
Invalid range requested.
12
The upper threshold must be greater than or equal to the lower
threshold.
Invalid upper threshold
requested
13
The I2C bus has failed on this analog card.
Analog card I2C failure
14
The 10mA Ac and DC probe excitation currents are not within
tolerance.
10mA AC/DC Ic out of range
15
The 6.3mA Ac and DC probe excitation current are not within
tolerance.
6.3mA AC/DC Ic out of range
16
The 0.63mA AC and DC probe excitation currents are not within
tolerance.
0.63mA AC/DC Ic out of range
17
The 100mA AC probe excitation current is not within tolerance.
100mA AC Ic out of range
18
The 10mA AC probe excitation current is not within tolerance.
10mA AC Ic out of range
19
The 6.3mA AC probe excitation current is not within tolerance.
6.3mA AC Ic out of range
20
The 0.63mA AC probe excitation current is not within tolerance.
0.63mA AC Ic out of range
21
The 100mA DC probe excitation current is not within tolerance.
100mA DC Ic out of range
22
The 10mA DC probe excitation current is not within tolerance.
10mA DC Ic out of range
23
The 6.3mA DC probe excitation current is not within tolerance.
6.3mA DC Ic out of range
24
The 0.63mA DC probe excitation current is not within tolerance.
0.63mA DC Ic out of range
25
The 2.5V reference is not within tolerance.
2.5V ref out of tolerance
26
The zero crossing divider output is not within tolerance.
Zero cross out of tolerance
27
IC Ref Test Failed for 100mA AC probe Current.
IC Ref Test Failed 100mA AC
28
IC Ref Test Failed for 100mA DC probe Current.
IC Ref Test Failed 100mA DC
29
IC Ref Test Failed for 100mA AC probe Current.
IC Ref Test Failed 10mA AC
30
IC Ref Test Failed for 100mA DC probe Current.
IC Ref Test Failed 10mA AC
31
Main ADC Ground Test Failed.
Main ADC Ground Test Failed