Section 6 – Flux Density Measurement
6-7
Table 6-3 AC Mode Analog Filter Selections
Filter Selection
Bandwidth
Full Bandwidth
1 Hz to 50 kHz
High Frequency
20 Hz to 50 kHz
Medium Frequency
20 Hz to 10 kHz
Low Frequency
20 Hz to 500 Hz
X Low Frequency
1 Hz to 500 Hz
No Corrections
1Hz to 50 kHz
Setting the Analog Filters
To select the analog filter:
1) Navigate the menu system to select the desired channel, then choose
AC Freq Setup and select the appropriate filter.
Figure 6-6 Analog Filter Selection
NOTE:
When operating the meter in Waveform display mode all signals will be
displayed on a common time base and the setting of the AC Freq Setup for All
Channels in the Waveform menu will over-ride and change any previous
individual channel AC Freq Setup settings.