F.W. BELL 8000 Series Gauss/Tesla Meter Instruction Manual
Section 6 – Flux Density Measurement
6-5
Figure 6-4 Measurement Mode Selection
DC Mode
Operation
In the DC mode, the presence of an AC component can lead to unstable
readings. If the peak value of the combined AC and DC component reaches
the electrical limits of the meter, even though the average DC level is within the
limits, an over range condition will appear on the display. This situation can
also lead to erratic behavior if the automatic ranging feature is being used.
The presence of an AC signal can be verified by observing the analog output
signal or by using the AC mode to determine the magnitude of the AC
component. Analog outputs are discussed later in this section.
AC Mode
Operation
It is possible for the flux density signal to contain both a DC component and
an AC component. In the AC mode the value displayed is the true RMS
value of the waveform with its DC component removed. If the DC
component is very large compared to the AC component, up to 1.25% of its
value may be seen as an error signal combined with the AC value.
Frequency Measurement
When used in the AC mode, the meter will display the frequency or period of
an AC field on the frequency indicator line, see Figure 6-5.
Signal: 5000.0 HZ
Frequency Reading
Units(Frequency / Period)
Figure 6-5 Frequency / Period Indicator
Note: Depending on the selection in the UNITS menu, frequency or period
may be displayed.