F.W. BELL 8000 Series Gauss/Tesla Meter Instruction Manual
Section 8 – Remote Operation
8-23
SCPI Message
Terminators
When transmitting a string to the instrument the message must be
“terminated” properly to notify the instrument that the message is
complete. One way is to append an ASCII line feed (LF) character as the
final character in the string, which is a 0A hex or 00001010 binary. Note
that 0A hex is equivalent to 10 decimal, but sending the two ASCII
characters “10” will not work. It must be the single byte representation of
the LF control character.
When using the RS-232 or the Ethernet port the instrument will always
send the LF (line feed) and CR (carriage return) characters every time it
transmits a message to the system controller.
Carriage Return >> ASCII code 13
Line Feed >> ASCII code 10
SCPI Meter
Specific
Commands
In the following tables and descriptions the commands are written such
that the short form of the command is written in UPPER CASE letters and
the remainder of the command is written in lower case letters. Either form
can be used.
If parameters are required they will appear within <> brackets. A <b>
parameter indicates a Boolean function, either “0” or “OFF”, or “1” or “ON”.
A <n> parameter is a single ASCII digit. <NRf> is usually a multiple digit
number and means Numeric Representation format.
Many commands are channel specific, so the “#” character in these
commands should be replaced with the channel number 1, 2, or 3 for
multi-channel instruments such as the 8030 model. If the channel number
is omitted then Channel-1 is assumed.
The gauss / tesla meter supports a subset of the available SCPI
commands. However, there are some functions that are not supported
with standard SCPI commands. In these cases these special commands
are patterned after other SCPI commands that are similar in function.