![F.W. Bell 8000 series Скачать руководство пользователя страница 108](http://html.mh-extra.com/html/f-w-bell/8000-series/8000-series_instruction-manual_535141108.webp)
F.W. BELL 8000 Series Gauss/Tesla Meter Instruction Manual
Section 8 – Remote Operation
8-29
The contents of the instrument’s real time calendar will be returned using
the following format:
<year>, <month>, <day>
where: <year> is in the range of 2000 – 9999
<month> is in the range of 1 – 12 (1 = January, 2 = February, etc)
<day> is in the range of 1 – 31 (day of the month)
:SYSTem:NETWork:ADDRess?
This command returns the instrument’s network IP address.
:SYSTem:LOGGing:DATa:REMove
This command deletes all previously stored data log files from the
intrument’s internal memory.
Status Register
Commnads
The Status Register Commands control and query the MEASUREMENT
EVENT, OPERATION EVENT and QUESTIONABLE EVENT registers as
well as the error queue.
:STATus:MEASurement:EVENt?
:STATus:OPERation:EVENt?
:STATus:QUEStionable:EVENt?
These commands return the contents of the specified EVENT register and
then clear that event register contents.
:STATus:MEASurement:ENABle <NRf>
:STATus:OPERation:ENABle <NRf>
:STATus:QUEStionable:ENABle <NRf>
These commands program the specified EVENT ENABLE register with the
value <NRf>. <NRf> is an ASCII string representing an integer mask. For
instance a value of 45 decimal is the same as binary 00101101, thus
setting bits 5, 3, 2 and 0 in the enable register.
:STATus:MEASurement:ENABle?
:STATus:OPERation:ENABle?
:STATus:QUEStionable:ENABle?
These commands return the contents of the specified EVENT ENABLE
register.
:STATus:MEASurement:CONDition?
:STATus:OPERation:CONDition?
:STATus:QUEStionable:CONDition?
These commands return the contents of the specified EVENT CONDITION
register. The EVENT CONDITION register is a real-time register reflecting
the state of the instrument at the time of the read. Some conditions could
happen very quickly and could be missed by this query. It is often better to
rely on the contents of the EVENT register since it latches the event until
Содержание 8000 series