![F.W. Bell 8000 series Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/f-w-bell/8000-series/8000-series_instruction-manual_535141020.webp)
F.W. BELL 8000 Series Gauss/Tesla Meter Instruction Manual
Section 2 - Specifications
2-1
SECTION 2 Specifications
Meter Specifications
Full-scale ranges are shown in the Tables 2-1 through 2-3 below, listed by probe type. In all
cases, the display resolution is 1 part in 300,000.
Table 2-1 Ranges for Low Field Probes
gauss (G)
tesla (T)
oersted (Oe)
ampere-turn/meter (A/m)
300.000 mG
30.0000
µ
T
300.000 mOe
23.8732 A/m
3.00000 G
300.000
µ
T
3.00000 Oe
238.732 A/m
Notes:
•
Low Field MOX type probes have an upper calibrated range of
±
2G but will respond to flux
densities of
±
3G (300µT) with increased error.
•
MOS type probes have a measurement limit of
±
1G (100µT) on a single range.
•
Display resolution for MOS type probes is 10µG (1nT).
•
Display resolution for MOX type probes is 1µG (0.1nT) but useful resolution is limited to 5-10µG
(1nT) due to noise & drift.
Table 2-2 Ranges for Mid Field Probes
(1x type)
gauss (G)
tesla (T)
oersted (Oe)
ampere-turn/meter (A/m)
30.0000 G
3.00000 mT
30.0000 Oe
2.38732 kA/m
300.000 G
30.0000 mT
300.000 Oe
23.8732 kA/m
3.00000 kG
300.000 mT
3.00000 kOe
238.732 kA/m
30.0000 kG
3.00000 T
30.0000 kOe
2.38732 MA/m
Table 2-3 Ranges for High Field Probes
(10x type)
gauss (G)
tesla (T)
oersted (Oe)
ampere-turn/meter (A/m)
300.000 G
30.0000 mT
300.000 Oe
23.8732 kA/m
3.00000 kG
300.000 mT
3.00000 kOe
238.732 kA/m
30.0000 kG
3.00000 T
30.0000 kOe
2.38732 MA/m
300.000 kG
30.0000 T
300.000 kOe
23.8732 MA/m
Содержание 8000 series
Страница 1: ...8000 SERIES GAUSS TESLA METER Instruction Manual ...
Страница 2: ...THIS SIDE BLANK Inside of Front Cover ...
Страница 37: ...F W BELL 8000 Series Gauss Tesla Meter Instruction Manual Section 4 Setup 4 8 This Page Intentionally Blank ...
Страница 132: ...OECO LLC 4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www fwbell com ...