Measurement Ranges
Scaling the Measure Value
At the analog output of the ES430.1, the following dependencies on the output
voltage apply for the measure value output:
TIP
At the analog output, lambda measure values of max. lambda=10 can be dis
-
played; in the calibration software, lambda measure values of max.
lambda=16 can be output.
Measure Value
Min
Min [V] Max
Max [V]
Lambda
0.60
0.60
10.0
10.0
1 / Lambda
0.0625 0.3125 1.67
8.333
Air/fuel ratio, A/F
8.5
0.425
200
10.0
Fuel/air ratio, F/A
0.005
0.25
0.118
5.90
Oxygen content O
2
[%]
0.0
0.0
25.0
10.0
Pump current, lambda sensor I
P
[mA] -3.0
3.50
+6.0
8.00
Measure Value
Measure Value Scaling
Unit
Lambda
-
1 / Lambda
-
Air/fuel ratio, A/F
-
Fuel/air ratio, F/A
-
Oxygen content O
2
%
Pump current, lambda sensor I
P
mA
=
U
out
1 V
=
U
out
5 V
1
=
U
out
0.05 V
A
F
=
U
out
50 V
F
A
O
2
=
U
out
0.4 V
I
p
=
U
out
- 5 V
0.5 V