ETAS
Troubleshooting
ES430.1 - User’s Guide
71
Measurement
does not start.
Are you being prompted to
carry out an update in the
INCA-Monitorlog or in the
Config-Tool?
Update the modules.
Is there no data from the
module?
Check that your power supply and test
setup correspond to the requirements
listed in section 4.7 on page 30.
Check that the wiring of the hardware to
the PC is correct/intact.
Check that the modules in the module
chain are connected correctly.
You are using the ES4xx
Configuration Tool and the
module is supplying no
data?
Check whether the position of one or
more modules in the chain has changed.
Check that you are not using an incorrect
A2L file.
Check whether you
have loaded the measure configuration
to the module chain.
Check that you have not assigned the
same IP address to two module chains.
Is the module supplying
usable data?
Check that the sensor is connected cor
-
rectly.
Data is lost during
data transfer.
Are you using WLAN in your
test setup?
WLAN is not permissible within this ETAS
network. Wire your test setup (ETAS
modules and their connection to the PC)
with ETAS cables only.
Are you using the correct
type of network card in your
laptop?
Check whether you are using a PCMCIA
network card in your laptop. PCMCIA
cards with an 8- or 16-bit data bus are not
suitable. Only use PCMCIA cards with a
32-bit data bus, mini-PCI or Express
-
Cards.
Problem
Diagnostic Questions
Possible Solutions