11 • General Technical Data
8
0113-90
1
08
.1
1
CHAPTER 11
General Technical Data
Power connection
Rated supply voltage
100 V – 120 V ± 10% / 220 V – 240 V ± 10%
Rated supply frequency
50 / 60 Hz
Line current
8 A / 4 A
Power input in standby mode
40 watts
Power input with max. HF output
500 watts / 920 VA
Terminal for grounding (potential equalization)
yes
Power fuses
T 8 A / T 4 A
Operating mode
Intermittent operation
ON time 25% (e.g. activated for 10 sec. / deactivated for
30 sec.)
Dimensions and weight
Width x height x depth
410 x 165 x 380 mm
Weight
9.5 kg
Ambient conditions for transport and storage of unit
Temperature
-40 °C to + 70 °C
Relative humidity
10% – 95%
Ambient conditions for operation of unit
Temperature
+10 °C to + 40 °C
Relative humidity
15% – 80%, noncondensing
Acclimatizing
If the unit has been stored or transported at temperatures below +10 °C or above +40 °C, the unit will require approx.
3 hours to acclimatize at room temperature.
Содержание VIO 300 D
Страница 1: ...VIO 300 D Configured for da Vinci User Manual 08 11 V 2 5 x ERBE ISI PN 551150 01 Rev A English...
Страница 2: ......
Страница 3: ...VIO 300 D Configured for da Vinci User Manual...
Страница 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Страница 22: ...3 Accessories 22 54 80113 901 08 11...
Страница 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Страница 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Страница 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Страница 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Страница 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...