CAUTION
Very long activation cycles without cooling phases
The electrosurgical unit is designed and tested for a relative ON time
of 25 % (conforming to IEC 60601-2-2). If you perform very long ac-
tivation cycles without appropriate cooling phases, the unit can be
damaged.
¨
Keep to the 25 % relative ON time (see also Technical Data, Op-
erating Mode), if you operate the unit for a lengthy period.
Notes
Grounding
Note:
If necessary, the equipment can be connected to the external grounding sys-
tem of the room with the grounding pin on the back of the unit and/or Cart using a
connecting cable designed for this purpose. Affects of low frequency leakage cur-
rents due to a defective grounding system within the room may be eliminated
through external grounding.
Use of a defibrillator
Note:
The equipment conforms to the requirements of Type CF and is protected
against the effects of a defibrillator discharge.
Membrane keyboards
Note:
If alcohol-based disinfectants are used on units with membrane keyboards,
this remove the anti-glare finish. However, the user surfaces remain fully functional.
This does not present a hazard.
Содержание VIO 300 D
Страница 1: ...VIO 300 D Configured for da Vinci User Manual 08 11 V 2 5 x ERBE ISI PN 551150 01 Rev A English...
Страница 2: ......
Страница 3: ...VIO 300 D Configured for da Vinci User Manual...
Страница 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Страница 22: ...3 Accessories 22 54 80113 901 08 11...
Страница 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Страница 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Страница 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Страница 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Страница 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...