37
/ 54
8 • Installation
8
0113-90
1
08
.1
1
WARNING
Incorrect line fuse, defective device
Risk of electric shock to the patient and medical personnel! Risk of
damage to property.
¨
Blown line fuses may only be replaced by a competent techni-
cian. Only replacement fuses that have the same rating as the
one specified on the unit’s rating plate may be used.
¨
When a fuse has been changed, the function of the unit must be
verified. If the unit does not function properly or if there are any
concerns, please contact ERBE.
WARNING
Damaged device, damaged accessories, modified device,
and modified accessories
Risk of burns and injury to the patient and medical personnel! Risk
of damage to property.
¨
Check the unit and accessories for damage every time before
use.
¨
You must not use damaged equipment or damaged accessories.
Replace defective accessories.
¨
If the equipment or equipment cart is damaged, please contact
our customer service.
¨
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in-
validate liability on the part of ERBE Elektromedizin GmbH.
Installing the electrosurgical unit
Fig. 8-1
MF receptacle input
(1)
Connect the
VIO 300 D to ICB Cable
to the MF receptacle. Run the cable to the ICB
box.
Содержание VIO 300 D
Страница 1: ...VIO 300 D Configured for da Vinci User Manual 08 11 V 2 5 x ERBE ISI PN 551150 01 Rev A English...
Страница 2: ......
Страница 3: ...VIO 300 D Configured for da Vinci User Manual...
Страница 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Страница 22: ...3 Accessories 22 54 80113 901 08 11...
Страница 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Страница 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Страница 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Страница 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Страница 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...