CHAPTER 10
Status Messages, Error Messages
An error message consists of an error code and an error text. The display of the VIO
system shows two different types of error messages.
a)
Error messages that prompt you to take action and remedy the error. You will find
these error messages in the table.
b)
Error messages that prompt you to inform Technical Service. These error messages
are not listed individually in the User Manual because the error texts of the relevant
error codes are constantly repeated. The error texts are:
•
Activation has been stopped. Activate again. If the display shows this error num-
ber repeatedly, please inform Technical Service.
•
Minor deviation from the system parameters. If the display shows this advice
repeatedly, please inform Technical Service.
Status Messages
B-84
Connected two-pedal footswitch ready for operation.
B-85
Two-pedal footswitch disconnected from system.
B-88
Single-pedal footswitch ready for operation.
B-89
Single-pedal footswitch disconnected from system.
B-93
Multifunctional footswitch ready for operation..
B-94
Multifunctional footswitch disconnected from system.
B-95
Connected instrument ready for operation. It has already been used approx. xxx times.
B-A6
Data transmission. Transferring data to program memory. Please wait until system has been restarted.
B-9B
Remote control. VIO system disconnected from an external master unit and ready for operation.
B-9C
Remote control. VIO system disconnected from external master unit.
B-9D
Remote control. VIO system controlled by external remote control and ready for operation.
B-9E
Remote control. VIO system disconnected from external remote control.
B-9F
Instrument disconnected from VIO system.
Error Messages
B-B
Nessy contact. Please check contact between skin and neutral electrode (patient plate).
B-F
Keyboard fault. The selection buttons are defective. If this message reappears, please inform Technical Ser-
vice.
B-01
Fault. Restarting device due to fault.
B-09
Fault. Restarting device due to fault.
Содержание VIO 300 D
Страница 1: ...VIO 300 D Configured for da Vinci User Manual 08 11 V 2 5 x ERBE ISI PN 551150 01 Rev A English...
Страница 2: ......
Страница 3: ...VIO 300 D Configured for da Vinci User Manual...
Страница 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Страница 22: ...3 Accessories 22 54 80113 901 08 11...
Страница 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Страница 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Страница 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Страница 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Страница 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...