1 • Safety Instructions
16
/ 54
8
0113-90
1
08
.1
1
CAUTION
Interference with electronic equipment due to the electro-
surgical unit
The activated electrosurgical unit can affect the performance of
electronic equipment by causing interference.
The equipment may fail or not perform properly.
¨
Position the electrosurgical unit, the cords of the instruments,
and the cord of the return electrode as far away as possible from
electronic equipment.
¨
Position the cords as far away as possible from the cords of elec-
tronic equipment.
CAUTION
Use of non-approved internal cables by Technical Service
This can result in the increased emission of electromagnetic waves
or reduce the immunity of the device.
The unit may fail or not perform properly.
¨
Technical Service may only use the internal cables that are listed
in the service manual for the device.
CAUTION
Stacked devices
If you stack the device next to other equipment or with other equip-
ment, the devices can affect each other.
The unit may fail or not perform properly.
¨
The device may only be stacked next to or with VIO series units.
¨
If it is necessary to operate the device near other equipment or
stacked together with other equipment, check whether the de-
vices are affecting each other: Are the devices behaving unusu-
ally? Do errors occur?
Damage to the unit and accessories
CAUTION
Alcohol-based spray disinfectant for fast disinfection
With membrane keyboards and paint surfaces there is the risk of
cracks. Propanol and ethanol will erode surfaces.
¨
Do not use these substances.
CAUTION
Alternate use of disinfectant solutions based on different
active ingredients
A color reaction may occur with plastics.
¨
Do not use these substances alternately.
Содержание VIO 300 D
Страница 1: ...VIO 300 D Configured for da Vinci User Manual 08 11 V 2 5 x ERBE ISI PN 551150 01 Rev A English...
Страница 2: ......
Страница 3: ...VIO 300 D Configured for da Vinci User Manual...
Страница 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Страница 22: ...3 Accessories 22 54 80113 901 08 11...
Страница 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Страница 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Страница 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Страница 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Страница 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...