1 • Safety Instructions
8
0113-90
1
08
.1
1
WARNING
Undesirable contact between the active instrument and
metal objects in the patient's body
Contact with metal hemostats, etc.
Risk of burns to the patient!
¨
Do not touch metal objects (e.g. implants) in the patient's body
with the active instrument.
Defective unit
WARNING
Undesirable rise in output level due to failure of electrosurgi-
cal unit
Risk of accidental tissue damage to the patient!
¨
The device shuts off independently.
¨
To guard against a possible failure of the electrosurgical unit,
have the device checked for safety at least once a year.
WARNING
Technical safety checks not being done
Risk of injury or death for patients and medical staff! Risk of damage
to property.
¨
Have the device checked for safety at least once a year.
¨
You must not use a device that is not safe.
WARNING
Failure of display elements
If display elements fail, you can no longer operate the device safely.
Risk of injury or death for patients and medical staff!
¨
You must not use the unit.
Interference caused by the unit
WARNING
Interference with cardiac pacemakers, internal defibrillators,
or other active implants
Activation of the electrosurgical unit may affect the performance of
active implants or damage them.
Risk of injury or death for patients!
¨
In the case of patients having active implants, consult the man-
ufacturer of the implant or the competent department of your
hospital prior to performing surgery.
Содержание VIO 300 D
Страница 1: ...VIO 300 D Configured for da Vinci User Manual 08 11 V 2 5 x ERBE ISI PN 551150 01 Rev A English...
Страница 2: ......
Страница 3: ...VIO 300 D Configured for da Vinci User Manual...
Страница 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Страница 22: ...3 Accessories 22 54 80113 901 08 11...
Страница 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Страница 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Страница 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Страница 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Страница 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...