10 • Status Messages, Error Messages
42
/ 54
8
0113-90
1
08
.1
1
B-10
Please end activation! Activation via finger or footswitch must be ended. After which reactivation is possi-
ble.
B-12
Please end activation! Footswitch or fingerswitch activation detected during device start-up.
B-16
Program memory full. Please delete programs no longer needed.
B-17
Double activation. Two switches pressed simultaneously e.g. footswitch and fingerswitch.
B-19
Line voltage fault. The unit has discontinued activation due to an insufficient supply voltage. If this recurs,
please inform Technical Service.
B-21
Invalid BMP file. Inform Technical Service.
B-22
Please end activation! Please remove forceps from tissue After which reactivation is possible.
B-26
The maximum application time in PRECISE APC mode was exceeded. The maximum application time can
be adjusted in the "Choose Coag Effect" Effect submenu.
B-81
Invalid system component. The connected component is not compatible with the VIO system. Inform
Technical Service.
B-1B
Self-check active. Please wait until self-check is complete. The unit is then ready for use.
B-1C
ON time limitation. Maximum ON time exceeded. Maximum ON time can be adjusted in setup.
B-1D
Instrument detection fault Do not use instrument; have it checked.
B-1E
Pressed button detected. Button pressed on device during start-up. Release button. If fault cannot be
remedied, inform Technical Service.
B-1F
NESSY symmetry. When applying neutral electrode (patient plate), ensure that neutral electrode line of
symmetry runs towards the operating field.
B-8E
VIO socket 1 fault; restart VIO. If fault cannot be remedied, inform Technical Service.
B-8F
VIO socket 2 fault. Restart VIO. If fault cannot be remedied, inform Technical Service.
B-90
VIO socket 3 fault. Restart VIO. If fault cannot be remedied, inform Technical Service.
B-97
Program memory fault. Restoring basic program setting. If this recurs, please inform Technical Service.
B-98
Program memory fault. The stored program could not be called up. If this recurs, please inform Technical
Service.
B-99
Activation type unavailable. For further information, consult user manual.
B-9A
Please check time in system menu.
B-A0
No other mode can be selected for this instrument.
B-A3
Footswitch not assigned. Footswitch activated but not assigned to a socket.
B-A4
Two footswitches connected. Two footswitches of the same type connected. For further information,
consult user manual.
B-A8
Invalid system component. The connected component is not compatible with the VIO system. Inform
Technical Service.
B-A9
Please confirm settings. Cannot activate device until current settings have been confirmed.
B-AA
Cannot activate mode. Attempt made to activate a mode that is switched off or unavailable. For further
information, consult user manual.
B-AB
Instrument not connected. Socket activated to which no instrument is connected. Or attempt made to
activate an instrument with old, invalid software.
Error Messages
Содержание VIO 300 D
Страница 1: ...VIO 300 D Configured for da Vinci User Manual 08 11 V 2 5 x ERBE ISI PN 551150 01 Rev A English...
Страница 2: ......
Страница 3: ...VIO 300 D Configured for da Vinci User Manual...
Страница 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Страница 22: ...3 Accessories 22 54 80113 901 08 11...
Страница 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Страница 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Страница 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Страница 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Страница 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...