33
/ 54
7 • Bipolar Mode
8
0113-90
1
08
.1
1
CHAPTER 7
Bipolar Mode
BiClamp
Properties
Special COAG mode for ERBE BiClamp (bipolar clamp). With four effect graduations
you can adjust the coagulation performance exactly to the type of tissue involved.
The AUTO STOP function is adjusted to BiClamp and ends activation automatically
when the best coagulation effect is achieved.
Modulation
BiClamp is a modulated current waveform with alternating pulse and rest periods.
This ratio is set using "Modulation". This means the larger the "Modulation" value,
the longer the rest period is compared to the subsequent active current flow period.
Technical data
Type of HF voltage
modulated sinusoidal alternating volt-
age
Nominal frequency
350 kHz (at R
L
= 500 ohms) ± 10%
Crest factor
1.4 (at R
L
= 500 ohms)
Designed load resistance
25 ohms
Max. HF peak voltage
220 Vp
Number of effects
4
Consistency of effects
Automatic control of HF peak voltage
max. HF output
300 watts ± 20 %
Содержание VIO 300 D
Страница 1: ...VIO 300 D Configured for da Vinci User Manual 08 11 V 2 5 x ERBE ISI PN 551150 01 Rev A English...
Страница 2: ......
Страница 3: ...VIO 300 D Configured for da Vinci User Manual...
Страница 18: ...1 Safety Instructions 18 54 80113 901 08 11...
Страница 22: ...3 Accessories 22 54 80113 901 08 11...
Страница 26: ...4 Description of the Controls 26 54 80113 901 08 11...
Страница 30: ...5 Working with the Electrosurgical Unit 30 54 80113 901 08 11...
Страница 32: ...6 Description of receptacle hardware 32 54 80113 901 08 11...
Страница 44: ...10 Status Messages Error Messages 44 54 80113 901 08 11...
Страница 52: ...12 Information on electromagnetic compatibility EMC 52 54 80113 901 08 11...