Operation
Turbimax CUE21 / CUE22
20
Hauser
Function
Options
Info
Delay on (DLY
▴
)
1 to 30 s
Set the desired number of seconds for the "Delay
on" function.
Delay off (DLY
▾
)
1 to 30 s
Set the desired number of seconds for the "Delay
off" function.
8.1.3
Access code
Function
Options
Info
Alarm 1 (ALM1)
AUTO
CAL
CONFIG
On
Å
CODE
• On
• OFF
If the security feature is switched on to restrict
access, you are required to enter an access code if
you wish to switch from AUTO to a different
operating mode. The only valid code is 333. This
code cannot be changed.
8.1.4
Advanced settings
Function
Options
Info
Extended settings (EXTD)
AUTO
CAL
CONFIG
On
Å
EXTD
• On
• OFF
Switch on the "Extended settings" function to access
the configuration of the following options:
• Response speed
• Displayed resolution
• LCD backlight brightness
• Units
• Ultrasonic cleaning
• RS-485 parameters
• Desiccant alarm
Response speed
Function
Options
Info
Response speed (RESP)
AUTO
CAL
CONFIG
10
Å
RESP
1 to 100 %
Factory setting: 10
Select the desired response speed for the NTU
values that are displayed and output.
Select the lowest speed (highest number) to avoid
displaying the effects of air and other anomalies.
Select the highest speed if there are rapid changes
taking place in the process that require monitoring.
The displayed number is a relative speed. The
approximate response time (in seconds) is
calculated by multiplying the displayed number by
5.
Displayed resolution
Function
Options
Info
Resolution (RES)
AUTO
CAL
CONFIG
0.01
Å
RES
1 to 0.0001
Factory setting: 0.01
For readings below 10 NTU the device can display
values with a resolution of up to four digits to the
right of the decimal point.